Device recall: Siemens Artis Models: Interventional Fluoroscopic X-Ray Systems-Models: Artis zee floor- 10094135 Artis zee ceiling-…

Class II FDA enforcement event 94299. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message TUBE HOT, have a break . the system cannot be operated normally, may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Siemens Medical Solutions USA, Inc started this device recall on 21 Mar 2024. The firm address in the FDA record is in Malvern, PA. FDA put the recall in Class II. FDA event 94299 has 1 product record. The first FDA enforcement report date is 24 Apr 2024. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

US Nationwide distribution.

Recalling firm

Siemens Medical Solutions USA, Inc, Malvern, PA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 94299: product records
Recall numberProductQuantityClassStatusReport date
Z-1598-2024Siemens Artis Models: Interventional Fluoroscopic X-Ray Systems-Models: Artis zee floor- 10094135 Artis zee ceiling- 10094137 Artis zee multi-purpose- 10094139 Artis zee biplane- 10094141 Artis zeego- 10280959 Artis Q floor- 10848280 Artis Q ceiling- 10848281 Artis Q biplane- 10848282 Artis Q zeego- 10848283 Artis…583 unitsClass IIOngoing24 Apr 2024

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2024
970
Share of these events in Class II
87%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
400

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 94299. https://recallregister.com/fda/ii/94299/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: US Nationwide distribution.
What quantity of product is in FDA event 94299?
The FDA record gives this quantity: 583 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2024 have the same class?
87% of the 970 device events that started in 2024 are in Class II.