Device recall: ARTIS Pheno VE30A and VE40A, Model 10849000
Class II FDA enforcement event 98772. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Siemens Medical Solutions USA, Inc started this device recall on 12 Mar 2026. The firm address in the FDA record is in Malvern, PA. FDA put the recall in Class II. FDA event 98772 has 1 product record. The first FDA enforcement report date is 22 Apr 2026. The record gives the status Ongoing. 291 of the 710 device events that started in 2026 have more product records. The median device event of 2026 has 1 product record.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
U.S. and O.U.S.
Recalling firm
Siemens Medical Solutions USA, Inc, Malvern, PA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: FDA Mandated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1814-2026 | ARTIS Pheno VE30A and VE40A, Model 10849000 | 73 | Class II | Ongoing | 22 Apr 2026 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2026
- 710
- Share of these events in Class II
- 87%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 400
Other events of the firm
Official channels
- FDA enforcement report Z-1814-2026The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 98772. https://recallregister.com/fda/ii/98772/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 10 Nov 2026
- Newest FDA report date on the register
- 4 Nov 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
U.S. and O.U.S.
- What quantity of product is in FDA event 98772?
- The FDA record gives this quantity:
73
- How did the recall start?
- The FDA record says:
FDA Mandated
- How many device recalls of 2026 have the same class?