Recall Register

Device recall: ACUSON Juniper Diagnostic Ultrasound System, REF: 11335791, and ACUSON Juniper Select Diagnostic Ultrasound System…

Class II FDA enforcement event 94211. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

On ultrasound systems, when Cardiac DICOM SR feature is configured to display either minimum or maximum measured value, and multiple cardiac region measurements are made, and results are exported into the SR feature, then The SR viewer will display the LAST measured value, not min or max, which could contribute to patient condition misdiagnosis or negatively influence patient management decisions.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Siemens Medical Solutions USA, Inc. started this device recall on 23 Feb 2024. The firm address in the FDA record is in Issaquah, WA. FDA put the recall in Class II. FDA event 94211 has 3 product records. The first FDA enforcement report date is 1 May 2024. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution - US Nationwide including in the states of AL, AR, CA, CO , CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA , MD, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, PR, RI,…

Recalling firm

Siemens Medical Solutions USA, Inc., Issaquah, WA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

3 product records in the event.

FDA event 94211: product records
Recall numberProductQuantityClassStatusReport date
Z-1599-2024ACUSON Juniper Diagnostic Ultrasound System, REF: 11335791, and ACUSON Juniper Select Diagnostic Ultrasound System, REF: 11653093, running software version VA10x with DICOM SR Cardiac option, from VB10A to VB10G, VB11A, VB11B, VB30B and VB30C5090Class IIOngoing1 May 2024
Z-1600-2024ACUSON Redwood 1.0 Diagnostic Ultrasound System, REF: 11503314, running software version VA10A to VA10G628Class IIOngoing1 May 2024
Z-1601-2024ACUSON Maple Diagnostic Ultrasound System, REF: 11711750, ACUSON Maple Select, REF: 11711779, running software version VA10A67Class IIOngoing1 May 2024

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2024
970
Share of these events in Class II
87%
Product records in this event
3Median of the firm: 1
Events of the firm on the register
400

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 94211. https://recallregister.com/fda/ii/94211/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution - US Nationwide including in the states of AL, AR, CA, CO , CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA , MD, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, PR, RI,…
What quantity of product is in FDA event 94211?
The event has 3 product records. The first record gives this quantity: 5090
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2024 have the same class?
87% of the 970 device events that started in 2024 are in Class II.