Recall Register

Device recall: The following Pro.specta Emission Computed Tomography Systems that were manufactured with an SLD Look Ahead Optical…

Class II FDA enforcement event 92704. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

During a whole-body planar scan with auto-contour, there is a possibility the short-linear drive (SLD) look-ahead sensors may trigger earlier than expected and may result in detector 1 positioning further from the patient than needed leading to a reduction in image resolution.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Siemens Medical Solutions USA, Inc. started this device recall on 12 Jul 2023. The firm address in the FDA record is in Hoffman Estates, IL. FDA put the recall in Class II. FDA event 92704 has 1 product record. The first FDA enforcement report date is 9 Aug 2023. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Distribution was made to ID, KS, MA, NH, OH, TX, and VA. There was no government/military distribution. Foreign distribution was made to Argentina, Australia, Austria, Belgium, Canada, Colombia,…

Recalling firm

Siemens Medical Solutions USA, Inc., Hoffman Estates, IL.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 92704: product records
Recall numberProductQuantityClassStatusReport date
Z-2256-2023The following Pro.specta Emission Computed Tomography Systems that were manufactured with an SLD Look Ahead Optical Sensor Assembly: (a) Symbia Pro.specta Q3, Model 11364751; (b) Symbia Pro.specta X3, Model 11364752; (c) Symbia Pro.specta X7, Model 11364753.108 systemsClass IIOngoing9 Aug 2023

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2023
917
Share of these events in Class II
88%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
400

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 92704. https://recallregister.com/fda/ii/92704/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Distribution was made to ID, KS, MA, NH, OH, TX, and VA. There was no government/military distribution. Foreign distribution was made to Argentina, Australia, Austria, Belgium, Canada, Colombia,…
What quantity of product is in FDA event 92704?
The FDA record gives this quantity: 108 systems
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2023 have the same class?