Device recall: syngo.via RT Image Suite with syngo.via VB30, VB40, VB50 and VB60- Intended for Radiation Therapy Treatment Material…
Class II FDA enforcement event 88588. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
Potential for data loss when using the contouring or patient marking workflow when editing a structure set after saving and reopening a study, resulting in an incomplete structure set sent to the treatment planning system (TPS).
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Siemens Medical Solutions USA, Inc started this device recall on 16 Aug 2021. The firm address in the FDA record is in Malvern, PA. FDA put the recall in Class II. FDA event 88588 has 3 product records. The first FDA enforcement report date is 22 Sep 2021. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
US Nationwide distribution.
Recalling firm
Siemens Medical Solutions USA, Inc, Malvern, PA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
3 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2445-2021 | syngo.via RT Image Suite with syngo.via VB30, VB40, VB50 and VB60- Intended for Radiation Therapy Treatment Material Number: 10496180 | 126 units | Class II | Ongoing | 22 Sep 2021 |
| Z-2446-2021 | syngo.CT VA20A, VA30A and VA40A in: syngo.via RT Image Suite / Sim&Go SOMATOM go.Now SOMATOM go.Up SOMATOM go.All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro and SOMATOM X.cite Material Numbers: 11061620 11061628 11061630 11061640 11061660 11061670 | 44 systems | Class II | Ongoing | 22 Sep 2021 |
| Z-2447-2021 | MR MAGNETOM Systems with RT Image Suite with syngo MR VA20A, VA30A and VA31A with license SW-Key CT_RT_IMA_SUITE_SYNTHETIC_CT Material Numbers: 11060815 | 2 systems | Class II | Ongoing | 22 Sep 2021 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2021
- 900
- Share of these events in Class II
- 89%
- Product records in this event
- 3Median of the firm: 1
- Events of the firm on the register
- 400
Other events of the firm
Official channels
- FDA enforcement report Z-2445-2021The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 88588. https://recallregister.com/fda/ii/88588/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
US Nationwide distribution.
- What quantity of product is in FDA event 88588?
- The event has 3 product records. The first record gives this quantity:
126 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2021 have the same class?
- 89% of the 900 device events that started in 2021 are in Class II.