Recall Register

Device recall: Ysio, model no. 10281013 Product Usage: The product is a stationary X-Ray system.

Class II FDA enforcement event 85029. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

The Ysio system may process incorrect internal values for the position of its ceiling tube stand. If in this situation the user activates longitudinal or transversal stand movement (by using the buttons at the tube head or the remote console), the system’s controls may (depending on the values being processed) initiate a “correction” of the tube stand position, resulting in an unexpected movement.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Siemens Medical Solutions USA, Inc started this device recall on 10 Feb 2020. The firm address in the FDA record is in Malvern, PA. FDA put the recall in Class II. FDA event 85029 has 1 product record. The first FDA enforcement report date is 25 Mar 2020. The record gives the status Terminated, with the termination date 31 Aug 2020.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

US Nationwide distribution in the states of AR, CA, DC, FL, IA, IL, IN, KS, KY, MA, MD, MI, MN, MO, MS, NC, NE, NH, NJ, NV, NY, OH, OR, PA, SC, TX, UT, VA and WI.

Recalling firm

Siemens Medical Solutions USA, Inc, Malvern, PA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 85029: product records
Recall numberProductQuantityClassStatusReport date
Z-1501-2020Ysio, model no. 10281013 Product Usage: The product is a stationary X-Ray system.81 US/ 308 worldwideClass IITerminated25 Mar 2020

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2020
916
Share of these events in Class II
93%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
400

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 85029. https://recallregister.com/fda/ii/85029/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: US Nationwide distribution in the states of AR, CA, DC, FL, IA, IL, IN, KS, KY, MA, MD, MI, MN, MO, MS, NC, NE, NH, NJ, NV, NY, OH, OR, PA, SC, TX, UT, VA and WI.
What quantity of product is in FDA event 85029?
The FDA record gives this quantity: 81 US/ 308 worldwide
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2020 have the same class?
93% of the 916 device events that started in 2020 are in Class II.