Device recall: The E.CAM gamma camera system Product Usage: The E.CAM gamma camera system is a SPECT system and measures the…

Class II FDA enforcement event 73504. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

We received a customer complaint that Siemens reported via the MedWatch process (1423253-2015-00002). The result of a Hazard Analysis and CAPA investigation, we have determined that for e.cam or Symbia E systems that although the existing user manual is correct, additional supplemental instructions, with the inclusion of visual aids, may further clarify and emphasize the proper collimator change

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Siemens Medical Solutions USA, Inc. started this device recall on 26 Feb 2016. The firm address in the FDA record is in Hoffman Estates, IL. FDA put the recall in Class II. FDA event 73504 has 2 product records. The first FDA enforcement report date is 6 Apr 2016. The record gives the status Terminated, with the termination date 23 Dec 2016.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution - US Nationwide in the states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY,…

Recalling firm

Siemens Medical Solutions USA, Inc., Hoffman Estates, IL.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

2 product records in the event.

FDA event 73504: product records
Recall numberProductQuantityClassStatusReport date
Z-1262-2016The E.CAM gamma camera system Product Usage: The E.CAM gamma camera system is a SPECT system and measures the distribution of radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body3090 unitsClass IITerminated6 Apr 2016
Z-1263-2016The Symbia gamma camera system Product Usage: The Symbia gamma camera system is a computed tomography x-ray system which is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This…726 unitsClass IITerminated6 Apr 2016

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2016
1,232
Share of these events in Class II
91%
Product records in this event
2Median of the firm: 1
Events of the firm on the register
400

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 73504. https://recallregister.com/fda/ii/73504/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution - US Nationwide in the states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY,…
What quantity of product is in FDA event 73504?
The event has 2 product records. The first record gives this quantity: 3090 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2016 have the same class?
91% of the 1,232 device events that started in 2016 are in Class II.