Recall Register

Device recall: Siemens Symbia S Series system, a diagnostic analog or digital detector based planar (2- dimensional) nuclear medicine…

Class II FDA enforcement event 71867. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Fuses with the improper amperage rating were installed in the Symbia S, Symbia T Series and Symbia lntevo systems. These fuses, in conjunction with a specific failure mode of the Power Controller assembly, could lead to the overheating of the assembly.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Siemens Medical Solutions USA, Inc. started this device recall on 31 Jul 2015. The firm address in the FDA record is in Hoffman Estates, IL. FDA put the recall in Class II. FDA event 71867 has 3 product records. The first FDA enforcement report date is 26 Aug 2015. The record gives the status Terminated, with the termination date 23 Dec 2016.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution - USA including AL, AK, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC,…

Recalling firm

Siemens Medical Solutions USA, Inc., Hoffman Estates, IL.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

3 product records in the event.

FDA event 71867: product records
Recall numberProductQuantityClassStatusReport date
Z-2435-2015Siemens Symbia S Series system, a diagnostic analog or digital detector based planar (2- dimensional) nuclear medicine (NM) imaging system installed and used in only one location within an imaging facility or in a mobile/ transportable imaging environment.1183 SystemsClass IITerminated26 Aug 2015
Z-2436-2015Siemens Symbia T Series, a radiological imaging system that is a combination of a single photon emission computed tomography (SPECT) camera system for nuclear medicine images, and a computed tomography (CT) camera system for x-ray images.1700Class IITerminated26 Aug 2015
Z-2437-2015Siemens Intevo Series system, a radiological imaging system that is a combination of a single photon emission computed tomography (SPECT) camera system for nuclear medicine images, and a computed tomography (CT) camera system for x-ray images.185Class IITerminated26 Aug 2015

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2015
1,153
Share of these events in Class II
91%
Product records in this event
3Median of the firm: 1
Events of the firm on the register
400

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 71867. https://recallregister.com/fda/ii/71867/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution - USA including AL, AK, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC,…
What quantity of product is in FDA event 71867?
The event has 3 product records. The first record gives this quantity: 1183 Systems
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2015 have the same class?