Food, drugs, devices › Siemens Medical Solutions USA, Inc

Device recall: SIEMENS Ysio Max; a radiographic system used in hospitals, clinics, and medical practices. Ysio Max enables radiographic

FDA enforcement event 71115 · Class II · initiated 13 Apr 2015 · status Terminated · terminated 20 Mar 2017

DeviceVoluntary: Firm initiated4 product recordsMalvern, PA

Siemens Medical Solutions USA, Inc (Malvern, PA) initiated this device recall on 13 Apr 2015; FDA lists 4 product records under event 71115, classified Class II.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 91% of the 1,153 device enforcement events initiated in 2015 carry the same class.

Status: Terminated, terminated 20 Mar 2017. Initiation: voluntary: firm initiated. Siemens Medical Solutions USA, Inc has 400 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Siemens Medical Solutions USA, Inc, Malvern, PA.

Reason for recall

"Study and all acquired images deleted when using systems with software version VE10E. Acquisitions have to be repeated. Connection with portable detectors (MAX wi-D, MAX mini) is sporadic in image recovery. Pressing undo button on the RAD subtask card or previous series button on the Image subtask card during image readout. Sporadically, during an automatic or a manual RIS update."

Distribution

"Nationwide Distribution."

Product records

Recall numberProductQuantityClassStatusReport date
Z-1667-2015SIEMENS Ysio Max; a radiographic system used in hospitals, clinics, and medical practices. Ysio Max enables radiographic and tomographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients. It can also be used for intravenous, small…133 totalClass IITerminated3 Jun 2015
Z-1668-2015SIEMENS Axiom Luminos dRF Max, a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. The Axiom Luminos dRF is a device intended to visualize anatomical…133 totalClass IITerminated3 Jun 2015
Z-1669-2015SIEMENS Luminos Agile Max; a universal imaging system for radiographic and fluoroscopic studies. Using either film cassettes or a digital mobile flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. Luminos Agile is applicable to emergency treatment on an…133 totalClass IITerminated3 Jun 2015
Z-1670-2015SIEMENS Uroskop Omnia Max ; a solid state detector fluoroscopic X-Ray system, primarily for urological applications (functional x-ray diagnostic, endourology and minimal invasive urology/surgery). The system,, which includes a radiologic/urologic treatment table, may be used for urological, gastroenterological and…133 totalClass IITerminated3 Jun 2015

Context

91%of 2015 device events are Class II1,153 events that year
4product records in this eventfirm median: 1
400events on this register for the firm

Other enforcement events for Siemens Medical Solutions USA, Inc

InitiatedProduct (first record)TypeClassRecords
8 Jun 2026ARTIS icono ceiling. Model Number: 11328100.DeviceClass II1
20 May 2026ARTIS icono biplane. Model Number: 11327600.DeviceClass II3
12 Mar 2026ARTIS Pheno VE30A and VE40A, Model 10849000DeviceClass II1
10 Mar 2026ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700.DeviceClass II1
9 Mar 2026ARTIS pheno, Model 10849000. interventional fluoroscopic x-ray systemDeviceClass II2
29 Dec 2025Interventional Fluoroscopic X-Ray System: LUMINOS Q.namix R; Siemens Material Number (SMN): 11574002; With Software VersDeviceClass II2
19 Dec 2025SOMATOM X.ceed Software applications: syngo.CT Brain Quantification syngo.CT Vessel Hyperdensities syngo.CT ASPECTSDeviceClass II4
5 Dec 2025LUMINOS Lotus Max, Model 11574100 and Luminos dRF Max, Model 10762471DeviceClass II1
24 Oct 2025MAMMOMAT Revelation;DeviceClass II3
1 Oct 2025MAGNETOM Cima.X Upgrade. Model Number: 11689304.DeviceClass I1

All 400 events for Siemens Medical Solutions USA, Inc

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Study and all acquired images deleted when using systems with software version VE10E. Acquisitions have to be repeated. Connection with portable detectors (MAX wi-D, MAX mini) is sporadic in image recovery. Pressing undo button on the RAD subtask card or previous series button on the Image subtask card during image readout. Sporadically, during an automatic or a manual RIS update."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Nationwide Distribution."

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 20 Mar 2017. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.