Device recall: Siemens brand ACUSON SC2000 Ultrasound System, Model 10433816, hardware that contains version 4 of the MPI board…

Class II FDA enforcement event 64855. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

When using the V5M transducer, the display of the temperature on the ACUSON SC2000 system can be lower than the actual temperature of the V5M transducer. This occurs only if the system hardware contains revision 4 of the MPI Board. There is a potential to cause esophageal burns in patients.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Siemens Medical Solutions USA, Inc. started this device recall on 28 Mar 2013. The firm address in the FDA record is in Mountain View, CA. FDA put the recall in Class II. FDA event 64855 has 1 product record. The first FDA enforcement report date is 17 Apr 2013. The record gives the status Terminated, with the termination date 14 Feb 2014.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution - USA and countries of Australia, Austria, Bosnia Herzegovina, Belgium, Finland, France, Germany, India, Italy, Japan, Kuwait, Malaysia, Netherlands, Norway, P.R. China,…

Recalling firm

Siemens Medical Solutions USA, Inc., Mountain View, CA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 64855: product records
Recall numberProductQuantityClassStatusReport date
Z-1081-2013Siemens brand ACUSON SC2000 Ultrasound System, Model 10433816, hardware that contains version 4 of the MPI board 10434200; The SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neo-natal and Fetal Cardiac, Pediatric, Transespohageal, Adult Cephalic, Peripheral Vessel, Abdominal,…120 unitsClass IITerminated17 Apr 2013

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2013
1,236
Share of these events in Class II
90%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
400

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 64855. https://recallregister.com/fda/ii/64855/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution - USA and countries of Australia, Austria, Bosnia Herzegovina, Belgium, Finland, France, Germany, India, Italy, Japan, Kuwait, Malaysia, Netherlands, Norway, P.R. China,…
What quantity of product is in FDA event 64855?
The FDA record gives this quantity: 120 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2013 have the same class?
90% of the 1,236 device events that started in 2013 are in Class II.