Recall Register

Food, drugs, devices › Siemens Medical Solutions USA, Inc

Device recall: Artis icono biplane -angiography systems developed for single and biplane diagnostic imaging and interventional procedur

FDA enforcement event 90184 · Class II · initiated 11 Apr 2022 · status Ongoing

DeviceVoluntary: Firm initiated2 product recordsMalvern, PA

FDA enforcement event 90184 is a device recall by Siemens Medical Solutions USA, Inc of Malvern, PA, initiated 11 Apr 2022 and classified Class II, with 2 product records.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 88% of the 839 device enforcement events initiated in 2022 carry the same class.

Status: Ongoing. Initiation: voluntary: firm initiated. Siemens Medical Solutions USA, Inc has 400 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Siemens Medical Solutions USA, Inc, Malvern, PA.

Reason for recall

"if the protective earth is interrupted and an additional fault leads to a fault current, the detection of the fault current may not be ensured may cause an electric shock if a person is in contact with metallic parts of the Carm"

Distribution

"US Nationwide distribution."

Product records

Recall numberProductQuantityClassStatusReport date
Z-1245-2022Artis icono biplane -angiography systems developed for single and biplane diagnostic imaging and interventional procedures SMN: 1132760051 unitsClass IIOngoing15 Jun 2022
Z-1246-2022Artis icono floor-angiography systems developed for single and biplane diagnostic imaging and interventional procedures SMN: 1132770013 unitsClass IIOngoing15 Jun 2022

Context

88%of 2022 device events are Class II839 events that year
2product records in this eventfirm median: 1
400events on this register for the firm

Other enforcement events for Siemens Medical Solutions USA, Inc

InitiatedProduct (first record)TypeClassRecords
8 Jun 2026ARTIS icono ceiling. Model Number: 11328100.DeviceClass II1
20 May 2026ARTIS icono biplane. Model Number: 11327600.DeviceClass II3
12 Mar 2026ARTIS Pheno VE30A and VE40A, Model 10849000DeviceClass II1
10 Mar 2026ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700.DeviceClass II1
9 Mar 2026ARTIS pheno, Model 10849000. interventional fluoroscopic x-ray systemDeviceClass II2
29 Dec 2025Interventional Fluoroscopic X-Ray System: LUMINOS Q.namix R; Siemens Material Number (SMN): 11574002; With Software VersDeviceClass II2
19 Dec 2025SOMATOM X.ceed Software applications: syngo.CT Brain Quantification syngo.CT Vessel Hyperdensities syngo.CT ASPECTSDeviceClass II4
5 Dec 2025LUMINOS Lotus Max, Model 11574100 and Luminos dRF Max, Model 10762471DeviceClass II1
24 Oct 2025MAMMOMAT Revelation;DeviceClass II3
1 Oct 2025MAGNETOM Cima.X Upgrade. Model Number: 11689304.DeviceClass I1

All 400 events for Siemens Medical Solutions USA, Inc

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "if the protective earth is interrupted and an additional fault leads to a fault current, the detection of the fault current may not be ensured may cause an electric shock if a person is in contact with metallic parts of the Carm"

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "US Nationwide distribution."

Is the recall still ongoing?

FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.