Device recall: Transducer S7-3t UDI (01)00884838061439 REF 989605406772 Can be used with: CX50, EPIQ 7, Affiniti 50 and Affiniti 70…

Class II FDA enforcement event 84902. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of the transducer. This issue may result in affected transducers not initiating auto-cool during use which could result in patient injury

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Philips Ultrasound Inc started this device recall on 30 Aug 2019. The firm address in the FDA record is in Bothell, WA. FDA put the recall in Class II. FDA event 84902 has 4 product records. The first FDA enforcement report date is 13 May 2020. The record gives the status Terminated, with the termination date 27 Nov 2024.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution - US Nationwide distribution.

Recalling firm

Philips Ultrasound Inc, Bothell, WA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

4 product records in the event.

FDA event 84902: product records
Recall numberProductQuantityClassStatusReport date
Z-1841-2020Transducer S7-3t UDI (01)00884838061439 REF 989605406772 Can be used with: CX50, EPIQ 7, Affiniti 50 and Affiniti 70 Ultrasound systems - Product Usage: Diagnostic ultrasound imaging or fluid flow analysis of the human body - particularly transesophageal echocardiography (TEE) studiesN/AClass IITerminated13 May 2020
Z-1842-2020Transducer S7-3t UDI (01)00884838067530 REF 989605439232 Used as accessory with: iE33, HD11 Ultrasound Systems - Product Usage: Diagnostic ultrasound imaging or fluid flow analysis of the human body - particularly transesophageal echocardiography (TEE) studiesN/AClass IITerminated13 May 2020
Z-1843-2020Transducer S8-3t UDI (01)00884838073524 REF 989605431171 Can be used as accessory with: EPIQ 7, Affiniti 70, EPIQ 5, EPIQ CVxi Ultrasound systems - Product Usage: Diagnostic ultrasound imaging or fluid flow analysis of the human body - particularly transesophageal echocardiography (TEE) studiesN/AClass IITerminated13 May 2020
Z-1844-2020Transducer S8-3t UDI (01)00884838067523 REF 989605420183 Can be used with: iE33 Utrasound System - Product Usage: Diagnostic ultrasound imaging or fluid flow analysis of the human body - particularly transesophageal echocardiography (TEE) studiesN/AClass IITerminated13 May 2020

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2019
968
Share of these events in Class II
88%
Product records in this event
4Median of the firm: 2
Events of the firm on the register
22

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 84902. https://recallregister.com/fda/ii/84902/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution - US Nationwide distribution.
What quantity of product is in FDA event 84902?
The event has 4 product records. The first record gives this quantity: N/A
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2019 have the same class?
88% of the 968 device events that started in 2019 are in Class II.