Recall Register

Device recall: EPIQ Ultrasound Systems, Model: EPIQ Elite, EPIQ 5G, EPIQ 5C, EPIQ 5W, EPIQ 7G, EPIQ 7C, EPIQ 7W, EPIQ CVx, EPIQ CVxi

Class II FDA enforcement event 88313. Recall start date: . Status in the FDA record: Completed.

Reason for recall, in the words of FDA

Ultrasound system control panel arm swivel lock could fail, which could cause the control panel and monitor to swivel freely while a user is positioning the monitor, or transporting/moving the system, which could lead to patient or user musculoskeletal injury.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Philips Ultrasound Inc started this device recall on 22 Jul 2021. The firm address in the FDA record is in Bothell, WA. FDA put the recall in Class II. FDA event 88313 has 1 product record. The first FDA enforcement report date is 25 Aug 2021. The record gives the status Completed. 291 of the 900 device events that started in 2021 have more product records. The median device event of 2021 has 1 product record. The openFDA file has 900 device events that started in 2021 and 916 that started in 2020.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution - US Nationwide distribution including in the states of FL, UT, CA, CO, WA, MS, NC, NV, GA, LA, VA, SC, NE, TN, HI, OR, ID, MI, OH, IN, IL, OK, TX, PA, NJ, CT, NY, DE, SD, WI,…

Recalling firm

Philips Ultrasound Inc, Bothell, WA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 88313: product records
Recall numberProductQuantityClassStatusReport date
Z-2306-2021EPIQ Ultrasound Systems, Model: EPIQ Elite, EPIQ 5G, EPIQ 5C, EPIQ 5W, EPIQ 7G, EPIQ 7C, EPIQ 7W, EPIQ CVx, EPIQ CVxi8935Class IICompleted25 Aug 2021

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2021
900
Share of these events in Class II
89%
Product records in this event
1Median of the firm: 2
Events of the firm on the register
22

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 88313. https://recallregister.com/fda/ii/88313/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
10 Nov 2026
Newest FDA report date on the register
4 Nov 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution - US Nationwide distribution including in the states of FL, UT, CA, CO, WA, MS, NC, NV, GA, LA, VA, SC, NE, TN, HI, OR, ID, MI, OH, IN, IL, OK, TX, PA, NJ, CT, NY, DE, SD, WI,…
What quantity of product is in FDA event 88313?
The FDA record gives this quantity: 8935
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2021 have the same class?
89% of the 900 device events that started in 2021 are in Class II.