Recall Register

Food, drugs, devices › Philips Ultrasound, Inc.

Device recall: 5000 Compact Series Ultrasound Systems, Product/REF: 5500 CV/795141, 5500 P/795143, 5500 W/795144, 5500 G/795145, and Pr

FDA enforcement event 94328 · Class II · initiated 23 Mar 2024 · status Ongoing

DeviceVoluntary: Firm initiated1 product recordBothell, WA

FDA enforcement event 94328 is a device recall by Philips Ultrasound, Inc. of Bothell, WA, initiated 23 Mar 2024 and classified Class II, with 1 product record.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 87% of the 970 device enforcement events initiated in 2024 carry the same class.

Status: Ongoing. Initiation: voluntary: firm initiated. Philips Ultrasound, Inc. has 22 enforcement events on this register, with a median of 2 product records per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Philips Ultrasound, Inc., Bothell, WA.

Reason for recall

"Ultrasound system with: 1) transesophageal echocardiography transducer (TEE) connected/not selected 2) 1, 2 non-TEE connected 3) Imaging with non-TEE 4) Selected/not selected imaging transducer removed 5) Transducer connected 6) Non-TEE imaging continues, Then though TEE is unselected, it may have power, no temperature monitoring/control, and may result in temperature above limit and patient burns"

Distribution

"Worldwide - US Nationwide distribution including in the states of CA, CO, CT, FL, GA, IL, IN, KS, KY, MA, MD, ME, MI, NC, NH, NM, NY, OH, OK, PA, PR, SC, TN, TX, UT, VA and the countries of…"

Product records

Recall numberProductQuantityClassStatusReport date
Z-1757-20245000 Compact Series Ultrasound Systems, Product/REF: 5500 CV/795141, 5500 P/795143, 5500 W/795144, 5500 G/795145, and Product/REF: 5300 W/795146, 5300 P/795147, 5300 G/795148, used with Deluxe and Premium Carts that come with multiport adapters.363Class IIOngoing15 May 2024

Context

87%of 2024 device events are Class II970 events that year
1product record in this eventfirm median: 2
22events on this register for the firm

Other enforcement events for Philips Ultrasound, Inc.

InitiatedProduct (first record)TypeClassRecords
5 Sep 2025Philips 3D6-2 Ultrasound TransducerDeviceClass III11
3 Jul 20253D6-2 Transducer ProbeDeviceClass III69
28 May 2025Philips 3D9-3v Transducer.DeviceClass II21
9 Oct 2023Philips EPIQ Diagnostic Ultrasound System Models: EPIQ 7G(model # 795200), EPIQ 7C (Model # 795201), EPIQ 7W (Model # 79DeviceClass II2
30 Jun 20233D9-3v Transducer, accessory to Philips Ultrasound Systems: EPIQ Elite; Affiniti 30, 50, and 70; ClearVue 850; HD15; iU2DeviceClass II1
21 Sep 2021EPIQ CVxi Diagnostic Ultrasound Systems, Product Code: 795232DeviceClass II1
19 Aug 2021Philips Ultrasound, Inc. Sparq, Diagnostic Ultrasound System, REF 795090, 100-240V -, 50/60 Hz, 500 VA, CE2797, Rx Only,DeviceClass II2
6 Aug 2021EPIQ Diagnostic Ultrasound Systems: Software Versions 5.0, 5.0.1 and 5.0.2 Models: EPIQ 5 - 795120 EPIQ 5C - 795205 EPIQDeviceClass II2
22 Jul 2021EPIQ Ultrasound Systems, Model: EPIQ Elite, EPIQ 5G, EPIQ 5C, EPIQ 5W, EPIQ 7G, EPIQ 7C, EPIQ 7W, EPIQ CVx, EPIQ CVxiDeviceClass II1
24 Dec 2020Philips EPIQ Diagnostic Ultrasound System, Models EPIQ Elite, EPIQ 5G, EPIQ 5C, EPIQ 5W, EPIQ 7G, EPIQ 7C, EPIQ 7W, EPIQDeviceClass II1

All 22 events for Philips Ultrasound, Inc.

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Ultrasound system with: 1) transesophageal echocardiography transducer (TEE) connected/not selected 2) 1, 2 non-TEE connected 3) Imaging with non-TEE 4) Selected/not selected imaging transducer removed 5) Transducer connected 6) Non-TEE imaging continues, Then though TEE is unselected, it may have power, no temperature monitoring/control, and may result in temperature above limit and patient burns"

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide - US Nationwide distribution including in the states of CA, CO, CT, FL, GA, IL, IN, KS, KY, MA, MD, ME, MI, NC, NH, NM, NY, OH, OK, PA, PR, SC, TN, TX, UT, VA and the countries of…"

Is the recall still ongoing?

FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.