Recall Register

Food, drugs, devices › Siemens Medical Solutions USA, Inc

Device recall: AXIOM Luminos dRFsystem The Axiom Luminos dRF is intended to be used as a universal diagnostic imaging system for radiog

FDA enforcement event 68110 · Class II · initiated 11 Apr 2014 · status Terminated · terminated 15 Dec 2014

DeviceVoluntary: Firm initiated1 product recordMalvern, PA

FDA enforcement event 68110 is a device recall by Siemens Medical Solutions USA, Inc of Malvern, PA, initiated 11 Apr 2014 and classified Class II, with 1 product record.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 91% of the 1,176 device enforcement events initiated in 2014 carry the same class.

Status: Terminated, terminated 15 Dec 2014. Initiation: voluntary: firm initiated. Siemens Medical Solutions USA, Inc has 400 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Siemens Medical Solutions USA, Inc, Malvern, PA.

Reason for recall

"It was determined that under rare environmental conditions (i.e., extremely dry air in combination with a highly isolated floor) electrostatic discharges (ESD) >8 kV may trigger an unintended movement of the AXIOM Luminos dRFsystem which may lead to an emergency situation and pose danger to the patients, operating personnel or to the unit."

Distribution

"Class II Recall - US Distribution including the states of IL,VA, TN, PA, TX, MN, GA, MI, IN, OH, FL, CA, OR, NY, ME, MO, OK, CO and NJ."

Product records

Recall numberProductQuantityClassStatusReport date
Z-1711-2014AXIOM Luminos dRFsystem The Axiom Luminos dRF is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. The Axiom Luminos dRF is a…37Class IITerminated11 Jun 2014

Context

91%of 2014 device events are Class II1,176 events that year
1product record in this eventfirm median: 1
400events on this register for the firm

Other enforcement events for Siemens Medical Solutions USA, Inc

InitiatedProduct (first record)TypeClassRecords
8 Jun 2026ARTIS icono ceiling. Model Number: 11328100.DeviceClass II1
20 May 2026ARTIS icono biplane. Model Number: 11327600.DeviceClass II3
12 Mar 2026ARTIS Pheno VE30A and VE40A, Model 10849000DeviceClass II1
10 Mar 2026ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700.DeviceClass II1
9 Mar 2026ARTIS pheno, Model 10849000. interventional fluoroscopic x-ray systemDeviceClass II2
29 Dec 2025Interventional Fluoroscopic X-Ray System: LUMINOS Q.namix R; Siemens Material Number (SMN): 11574002; With Software VersDeviceClass II2
19 Dec 2025SOMATOM X.ceed Software applications: syngo.CT Brain Quantification syngo.CT Vessel Hyperdensities syngo.CT ASPECTSDeviceClass II4
5 Dec 2025LUMINOS Lotus Max, Model 11574100 and Luminos dRF Max, Model 10762471DeviceClass II1
24 Oct 2025MAMMOMAT Revelation;DeviceClass II3
1 Oct 2025MAGNETOM Cima.X Upgrade. Model Number: 11689304.DeviceClass I1

All 400 events for Siemens Medical Solutions USA, Inc

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "It was determined that under rare environmental conditions (i.e., extremely dry air in combination with a highly isolated floor) electrostatic discharges (ESD) >8 kV may trigger an unintended movement of the AXIOM Luminos dRFsystem which may lead to an emergency situation and pose danger to the patients, operating personnel or to the unit."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Class II Recall - US Distribution including the states of IL,VA, TN, PA, TX, MN, GA, MI, IN, OH, FL, CA, OR, NY, ME, MO, OK, CO and NJ."

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 15 Dec 2014. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.