Recall Register

Food, drugs, devices › Hill-Rom, Inc.

Device recall: Liko Overhead Rail System, Ceiling Mounted Rail Systems for all Liko Stationary Lifts including Likorall, Multirall and

FDA enforcement event 62334 · Class II · initiated 31 Oct 2012 · status Terminated · terminated 9 Mar 2016

DeviceVoluntary: Firm initiated1 product recordBatesville, IN

FDA enforcement event 62334 is a device recall by Hill-Rom, Inc. of Batesville, IN, initiated 31 Oct 2012 and classified Class II, with 1 product record.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 88% of the 823 device enforcement events initiated in 2012 carry the same class.

Status: Terminated, terminated 9 Mar 2016. Initiation: voluntary: firm initiated. Hill-Rom, Inc. has 28 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Hill-Rom, Inc., Batesville, IN.

Reason for recall

"Liko/Hill-Rom to date has received two reports from facilities alleging that the overhead rail system has failed to the extent that the overhead rails have fallen down while in use in high humidity (swimming pool) areas. Analysis of the incidents suggest that the components (hardware) used to secure the rail to the ceiling had corroded and may be unsuitable for environments with an atmosphere con"

Distribution

"Worldwide Distribution: USA (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV,…"

Product records

Recall numberProductQuantityClassStatusReport date
Z-0872-2013Liko Overhead Rail System, Ceiling Mounted Rail Systems for all Liko Stationary Lifts including Likorall, Multirall and Masterlift Systems Product Usage - The Overhead Rail System allows overhead lifts to be mounted to the ceiling allowing versatility for use in lifting situations. It can be used with different Liko…3816 in totalClass IITerminated27 Feb 2013

Context

88%of 2012 device events are Class II823 events that year
1product record in this eventfirm median: 1
28events on this register for the firm

Other enforcement events for Hill-Rom, Inc.

InitiatedProduct (first record)TypeClassRecords
14 Jan 2022(1)Traverse Rail Carrier E-System, wide 31017XX component to Liko overhead lift, (2)Traverse Rail Carrier 31025XX componDeviceClass II1
23 Dec 2021Liko Mobile Lifts: Primarily intended for use in nursing homes Liko M220, Model 2050010DeviceClass II2
14 Oct 2021These are communication systems, labeled as: a. NaviCare Nurse Call, Part Numbers: 3.6, 3.7, 3.9 b. Voalte Nurse Call, PDeviceClass II1
11 Aug 2021OctoStretch with Stretch Leveller - The combination of this products allows the user to easily adjust/control the centerDeviceClass II5
28 May 2021Affinity Four Birthing Bed-used as a birthing bed for women of child bearing age in an LDR (Labor, Delivery, Recovery) oDeviceClass II1
18 Dec 2020Liko Multirall 200 overhead lift- overhead lift is a general-purpose lift with the intended use in healthcare, intensiveDeviceClass I1
30 Jun 2020Progressa Bed, Catalog No. P7500 - Product Usage: intended to be used to treat or prevent pulmonary or other complicatioDeviceClass II2
30 Apr 2020Centrella Smart+ Bed, Catalog Number P7900DeviceClass II1
14 Feb 2020NaviCare Patient Safety Software Versions 3.5.400-3.9.300. The NaviCare Patient Safety (Patient Safety) Software is compDeviceClass II1
12 Jan 2018Progressa Bed - AC-powered adjustable hospital bedDeviceClass II1

All 28 events for Hill-Rom, Inc.

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Liko/Hill-Rom to date has received two reports from facilities alleging that the overhead rail system has failed to the extent that the overhead rails have fallen down while in use in high humidity (swimming pool) areas. Analysis of the incidents suggest that the components (hardware) used to secure the rail to the ceiling had corroded and may be unsuitable for environments with an atmosphere con"

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide Distribution: USA (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV,…"

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 9 Mar 2016. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.