Recall Register

Device recall: Liko Multirall 200 overhead lift- overhead lift is a general-purpose lift with the intended use in healthcare…

Class I FDA enforcement event 87053. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Multirall Q-link strap lock has not been fully engaged in the S65 rail carriage hook by user, leading to the Multirall lift becoming detached from the rail system, potentially resulting in the fall of both the lift motor and the patient

Class I, in the words of FDA

a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death

Hill-Rom, Inc. started this device recall on 18 Dec 2020. The firm address in the FDA record is in Batesville, IN. FDA put the recall in Class I. FDA event 87053 has 1 product record. The first FDA enforcement report date is 10 Feb 2021. The record gives the status Terminated, with the termination date 26 Sep 2023.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution: US (Nationwide) and countries of::AT, AU, BR, CA, CH, CZ, DE, DK, ES, FI, FR, GB, GR, IE, IL, IS, IT, JP, LV, NL, NO, NZ, SE & SK.

Recalling firm

Hill-Rom, Inc., Batesville, IN.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 87053: product records
Recall numberProductQuantityClassStatusReport date
Z-0925-2021Liko Multirall 200 overhead lift- overhead lift is a general-purpose lift with the intended use in healthcare, intensive care and rehabilitation. Affected devices: Liko Multirall 200 (product number 3130001) Universal SlingBar 450 R2R (product number 3156095) Universal SlingBar 350 R2R (product number 3156094)…10470 units Expanded Recall: 154 unitsClass ITerminated10 Feb 2021

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class I: The highest of the three FDA recall classes.

The event in context

Device events that started in 2020
916
Share of these events in Class I
4%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
28

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 87053. https://recallregister.com/fda/i/87053/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution: US (Nationwide) and countries of::AT, AU, BR, CA, CH, CZ, DE, DK, ES, FI, FR, GB, GR, IE, IL, IS, IT, JP, LV, NL, NO, NZ, SE & SK.
What quantity of product is in FDA event 87053?
The FDA record gives this quantity: 10470 units Expanded Recall: 154 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2020 have the same class?
4% of the 916 device events that started in 2020 are in Class I.