Recall Register

Food, drugs, devices › Hill-Rom, Inc.

Device recall: (1)Traverse Rail Carrier E-System, wide 31017XX component to Liko overhead lift, (2)Traverse Rail Carrier 31025XX compon

FDA enforcement event 89562 · Class II · initiated 14 Jan 2022 · status Terminated · terminated 7 Nov 2024

DeviceVoluntary: Firm initiated1 product recordBatesville, IN

FDA enforcement event 89562 is a device recall by Hill-Rom, Inc. of Batesville, IN, initiated 14 Jan 2022 and classified Class II, with 1 product record.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 88% of the 839 device enforcement events initiated in 2022 carry the same class.

Status: Terminated, terminated 7 Nov 2024. Initiation: voluntary: firm initiated. Hill-Rom, Inc. has 28 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Hill-Rom, Inc., Batesville, IN.

Reason for recall

"Traverse rail carriage delivered with non-conforming screws are too short to ensure appropriate installation. There is potential that the screws are unable to secure the traverse rail appropriately and could loosen over time causing the rail system to fall, potentially causing injury to caregivers and/or patients"

Distribution

"Worldwide distribution - US Nationwide and the countries of AT, CA, CZ, DE, DK, ES, FI, FR, GB, HK, IE, IL, IT, JP, NL, NO, NZ, RU, SE, SG, SK."

Product records

Recall numberProductQuantityClassStatusReport date
Z-0783-2022(1)Traverse Rail Carrier E-System, wide 31017XX component to Liko overhead lift, (2)Traverse Rail Carrier 31025XX component to Liko overhead lift. Intended for to make safe lifting of a patient possible. Affected Devices: 3101704 TRAV RAIL CARR E-SYST STD 3102514 TR CARRIER SLIM/RAISED, LR 3101705 TRAV RAIL CARR…Traverse Rail Carrier E-System, wide 31017XX 5 Traverse Rail Carrier 31025XX 1798Class IITerminated23 Mar 2022

Context

88%of 2022 device events are Class II839 events that year
1product record in this eventfirm median: 1
28events on this register for the firm

Other enforcement events for Hill-Rom, Inc.

InitiatedProduct (first record)TypeClassRecords
23 Dec 2021Liko Mobile Lifts: Primarily intended for use in nursing homes Liko M220, Model 2050010DeviceClass II2
14 Oct 2021These are communication systems, labeled as: a. NaviCare Nurse Call, Part Numbers: 3.6, 3.7, 3.9 b. Voalte Nurse Call, PDeviceClass II1
11 Aug 2021OctoStretch with Stretch Leveller - The combination of this products allows the user to easily adjust/control the centerDeviceClass II5
28 May 2021Affinity Four Birthing Bed-used as a birthing bed for women of child bearing age in an LDR (Labor, Delivery, Recovery) oDeviceClass II1
18 Dec 2020Liko Multirall 200 overhead lift- overhead lift is a general-purpose lift with the intended use in healthcare, intensiveDeviceClass I1
30 Jun 2020Progressa Bed, Catalog No. P7500 - Product Usage: intended to be used to treat or prevent pulmonary or other complicatioDeviceClass II2
30 Apr 2020Centrella Smart+ Bed, Catalog Number P7900DeviceClass II1
14 Feb 2020NaviCare Patient Safety Software Versions 3.5.400-3.9.300. The NaviCare Patient Safety (Patient Safety) Software is compDeviceClass II1
12 Jan 2018Progressa Bed - AC-powered adjustable hospital bedDeviceClass II1
23 Feb 2017TotalCare Bed Systems, Model Numbers P1900, PR1900, and U1900 with 3-Level Patient Position Monitoring and Scale UpgradeDeviceClass II1

All 28 events for Hill-Rom, Inc.

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Traverse rail carriage delivered with non-conforming screws are too short to ensure appropriate installation. There is potential that the screws are unable to secure the traverse rail appropriately and could loosen over time causing the rail system to fall, potentially causing injury to caregivers and/or patients"

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide distribution - US Nationwide and the countries of AT, CA, CZ, DE, DK, ES, FI, FR, GB, HK, IE, IL, IT, JP, NL, NO, NZ, RU, SE, SG, SK."

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 7 Nov 2024. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.