Device recall: Merlin@home Transmitter REF EX1100 The Merlin@home" (M@h) transmitter is intended to transfer stored data from the…

Class III FDA enforcement event 95557. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Due to heart transmitter not being able to upgrade to the current software version in the field after experiencing a power interruption during a prior over-the-wire software download.

Class III, in the words of FDA

a situation in which use of or exposure to a violative product is not likely to cause adverse health consequences

Abbott Medical started this device recall on 1 Oct 2024. The firm address in the FDA record is in Sylmar, CA. FDA put the recall in Class III. FDA event 95557 has 2 product records. The first FDA enforcement report date is 4 Dec 2024. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide - U.S. Nationwide distribution in the states of ALABAMA, ALASKA, ARKANSAS, CALIFORNIA, COLORADO, CONNECTICUT, DISTRICT OF COLUMBIA, FLORIDA, GEORGIA, ILLINOIS, INDIANA, KENTUCKY, MICHIGAN,…

Recalling firm

Abbott Medical, Sylmar, CA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

2 product records in the event.

FDA event 95557: product records
Recall numberProductQuantityClassStatusReport date
Z-0551-2025Merlin@home Transmitter REF EX1100 The Merlin@home" (M@h) transmitter is intended to transfer stored data from the patient s implanted heart device to Merlin.net" Patient Care Network (PCN) to facilitate remote monitoring by using inductive (EX1100 Model) or RF telemetry (EX1150 Model) to read device data and…130 transmittersClass IIIOngoing4 Dec 2024
Z-0552-2025Merlin@home with the MerlinOnDemand capability Transmitter REF EX1150 Merlin@home with the MerlinOnDemand" (MoD) capability is a transmitter configured for multiple patient use. It can receive information from an implantable heart device in unpaired mode. With MoD capability, the transmitter will retrieve full…PendingClass IIIOngoing4 Dec 2024

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class III: The lowest of the three FDA recall classes.

The event in context

Device events that started in 2024
970
Share of these events in Class III
2%
Product records in this event
2Median of the firm: 1
Events of the firm on the register
16

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 95557. https://recallregister.com/fda/iii/95557/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide - U.S. Nationwide distribution in the states of ALABAMA, ALASKA, ARKANSAS, CALIFORNIA, COLORADO, CONNECTICUT, DISTRICT OF COLUMBIA, FLORIDA, GEORGIA, ILLINOIS, INDIANA, KENTUCKY, MICHIGAN,…
What quantity of product is in FDA event 95557?
The event has 2 product records. The first record gives this quantity: 130 transmitters
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2024 have the same class?
2% of the 970 device events that started in 2024 are in Class III.