Recall Register

Device recall: Gallant VR Implantable Cardioverter Defibrillator: VR ICD REF CDVRA500Q Gallant DR Implantable Cardioverter…

Class II FDA enforcement event 92923. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Due to Bluetooth malfunction on a subset of Implantable Cardioverter Defibrillator (ICD) and Cardia Resynchronization Therapy Defibrillator *CRT-D) devices that were manufactured prior to April 2022 as a result of crystal component issue within the Bluetooth Low Energy (BLE) circuit. This results in the device entering an inductive telemetry only mode with loss of Bluetooth communication. In…

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Abbott Medical started this device recall on 16 Aug 2023. The firm address in the FDA record is in Sylmar, CA. FDA put the recall in Class II. FDA event 92923 has 2 product records. The first FDA enforcement report date is 27 Sep 2023. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

U.S.: Nation-wide O.U.S.: Australia, Austria, Bahrain, Belgium, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italy, Korea, Kuwait,…

Recalling firm

Abbott Medical, Sylmar, CA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

2 product records in the event.

FDA event 92923: product records
Recall numberProductQuantityClassStatusReport date
Z-2614-2023Gallant VR Implantable Cardioverter Defibrillator: VR ICD REF CDVRA500Q Gallant DR Implantable Cardioverter Defibrillator: DR ICD REF CDDRA500Q Neutrino VR Implantable Cardioverter Defibrillator: NxT VR ICD REF CDVRA600Q Neutrino DR Implantable Cardioverter Defibrillator: NxT DR ICD REF CDDRA600Q Entrant VR…38,033 devicesClass IIOngoing27 Sep 2023
Z-2615-2023Gallant HF Cardiac Resynchronization Therapy Defibrillator: HF CRT-D REF CDHFA500Q Neutrino HF Cardiac Resynchronization Therapy Defibrillator: NxT HF CRT-D REF CDHFA600Q Entrant HF Cardiac Resynchronization Therapy Defibrillator : HF CRT-D REF CDHFA300Q The primary function of ICDs and CRT-Ds are to treat…29,088 devicesClass IIOngoing27 Sep 2023

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2023
917
Share of these events in Class II
88%
Product records in this event
2Median of the firm: 1
Events of the firm on the register
16

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 92923. https://recallregister.com/fda/ii/92923/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: U.S.: Nation-wide O.U.S.: Australia, Austria, Bahrain, Belgium, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italy, Korea, Kuwait,…
What quantity of product is in FDA event 92923?
The event has 2 product records. The first record gives this quantity: 38,033 devices
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2023 have the same class?
88% of the 917 device events that started in 2023 are in Class II.