Device recall: Philips Patient Information Center iX configured with a regional setting which uses a comma for a decimal separator…
Class III FDA enforcement event 82914. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
Infusion Pump Data Storage Accuracy-Data from the BBraun Space LAN or Arcomed UniqueDoc infusion pumps transmitted via the HL7 output interface through the Patient Information Center iX using a LAN driver may be recorded in the patients chart or electronic medical record at exactly 100 times the actual bolus rate, infusion rate and total volume values
Class III, in the words of FDA
a situation in which use of or exposure to a violative product is not likely to cause adverse health consequences
Philips North America, LLC started this device recall on 29 Apr 2019. The firm address in the FDA record is in Andover, MA. FDA put the recall in Class III. FDA event 82914 has 2 product records. The first FDA enforcement report date is 19 Jun 2019. The record gives the status Terminated, with the termination date 22 Nov 2024.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide Distribution - US Nationwide Foreign: Andorra Aruba Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Brazil Bulgaria Canada Chile China Colombia Costa Rica Croatia…
Recalling firm
Philips North America, LLC, Andover, MA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
2 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1771-2019 | Philips Patient Information Center iX configured with a regional setting which uses a comma for a decimal separator: 866389 Patient Information Center iX B.01 (or later); 866390 Patient Information Center iX B.01 upgrade (or later) Product Usage: The intended use of the Information Center Software is to display… | 12,391 | Class III | Terminated | 19 Jun 2019 |
| Z-1772-2019 | Philips IntelliBridge System configured with a regional setting which uses a comma for a decimal separator: 866417 IntelliBridge System C.01 (or later) 866418 IntelliBridge System C.01 upgrade (or later) Product Usage: The IntelliBridge System is indicated for use in data collection and clinical information… | 180 units | Class III | Terminated | 19 Jun 2019 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class III: The lowest of the three FDA recall classes.
The event in context
- Device events that started in 2019
- 968
- Share of these events in Class III
- 6%
- Product records in this event
- 2Median of the firm: 1
- Events of the firm on the register
- 129
Other events of the firm
Official channels
- FDA enforcement report Z-1771-2019The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 82914. https://recallregister.com/fda/iii/82914/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide Distribution - US Nationwide Foreign: Andorra Aruba Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Brazil Bulgaria Canada Chile China Colombia Costa Rica Croatia…
- What quantity of product is in FDA event 82914?
- The event has 2 product records. The first record gives this quantity:
12,391
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2019 have the same class?
- 6% of the 968 device events that started in 2019 are in Class III.