Device recall: Brilliance iCT; Product Code (REF): 728306;

Class II FDA enforcement event 98248. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry…

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Philips North America Llc started this device recall on 7 Jan 2026. The firm address in the FDA record is in Cambridge, MA. FDA put the recall in Class II. FDA event 98248 has 4 product records. The first FDA enforcement report date is 18 Feb 2026. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide - US Nationwide distribution in the states of AL, AR, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MA, MD, MN, NC, NY, OH, OR, PA, TX, VT, WA, WV and the countries of Australia, Austria,…

Recalling firm

Philips North America Llc, Cambridge, MA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

4 product records in the event.

FDA event 98248: product records
Recall numberProductQuantityClassStatusReport date
Z-1315-2026Brilliance iCT; Product Code (REF): 728306;47 unitsClass IIOngoing18 Feb 2026
Z-1316-2026IQon Spectral CT; Product Code (REF): 728332;52 unitsClass IIOngoing18 Feb 2026
Z-1317-2026Spectral CT; Product Code (REF): 728333;283 unitsClass IIOngoing18 Feb 2026
Z-1318-2026Spectral CT 7500 on Rails; Product Code (REF): 728334;3 unitsClass IIOngoing18 Feb 2026

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2026
572
Share of these events in Class II
87%
Product records in this event
4Median of the firm: 1
Events of the firm on the register
129

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 98248. https://recallregister.com/fda/ii/98248/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide - US Nationwide distribution in the states of AL, AR, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MA, MD, MN, NC, NY, OH, OR, PA, TX, VT, WA, WV and the countries of Australia, Austria,…
What quantity of product is in FDA event 98248?
The event has 4 product records. The first record gives this quantity: 47 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2026 have the same class?
87% of the 572 device events that started in 2026 are in Class II.