Device recall: Medtronic Pulmonary Vein Ablation Catheter(R) GOLD (PVAC GOLD) 990078 - Product Usage: The Pulmonary Vein Ablation…
Class III FDA enforcement event 82319. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
The firm issued an updated instructions for use to reduce the potential for phrenic nerve injury.
Class III, in the words of FDA
a situation in which use of or exposure to a violative product is not likely to cause adverse health consequences
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) started this device recall on 4 Mar 2019. The firm address in the FDA record is in Mounds View, MN. FDA put the recall in Class III. FDA event 82319 has 1 product record. The first FDA enforcement report date is 3 Apr 2019. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
International distribution in the countries of Australia, Austria, Brazil, Canada, France, Germany, Hungary, Ireland, Israel, Italy, Latvia, Netherlands, Poland, Portugal, Spain, Switzerland, United…
Recalling firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), Mounds View, MN.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1069-2019 | Medtronic Pulmonary Vein Ablation Catheter(R) GOLD (PVAC GOLD) 990078 - Product Usage: The Pulmonary Vein Ablation Catheter GOLD is designed to be used for the creation of endocardial lesions (focal and linear) during cardiac ablation procedures for the treatment of symptomatic Atrial Fibrillation (AF). The PVAC GOLD… | 2917 units | Class III | Ongoing | 3 Apr 2019 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class III: The lowest of the three FDA recall classes.
The event in context
- Device events that started in 2019
- 968
- Share of these events in Class III
- 6%
- Product records in this event
- 1Median of the firm: 2
- Events of the firm on the register
- 30
Other events of the firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF): all 30 events
Official channels
- FDA enforcement report Z-1069-2019The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 82319. https://recallregister.com/fda/iii/82319/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
International distribution in the countries of Australia, Austria, Brazil, Canada, France, Germany, Hungary, Ireland, Israel, Italy, Latvia, Netherlands, Poland, Portugal, Spain, Switzerland, United…
- What quantity of product is in FDA event 82319?
- The FDA record gives this quantity:
2917 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2019 have the same class?
- 6% of the 968 device events that started in 2019 are in Class III.