Recall Register

Device recall: Medical Device Identification Cards associated with Sprint Quattro Lead, Product/CFN 6935M72

Class II FDA enforcement event 89054. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

The back of some Medical Device Identification Cards provided to patients were printed with an erroneous statement stating the following: "This patient has a complete MR conditional system implanted."

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) started this device recall on 8 Nov 2021. The firm address in the FDA record is in Mounds View, MN. FDA put the recall in Class II. FDA event 89054 has 2 product records. The first FDA enforcement report date is 22 Dec 2021. The record gives the status Terminated, with the termination date 4 May 2026.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

US, Canada, Northern Mariana Islands

Recalling firm

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), Mounds View, MN.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

2 product records in the event.

FDA event 89054: product records
Recall numberProductQuantityClassStatusReport date
Z-0390-2022Medical Device Identification Cards associated with Sprint Quattro Lead, Product/CFN 6935M72187 Medical Device CardsClass IITerminated22 Dec 2021
Z-0391-2022Medical Device Identification Cards associated with Astra XT DR Implantable pulse generator10 Medical Device CardsClass IITerminated22 Dec 2021

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2021
900
Share of these events in Class II
89%
Product records in this event
2Median of the firm: 2
Events of the firm on the register
30

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 89054. https://recallregister.com/fda/ii/89054/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: US, Canada, Northern Mariana Islands
What quantity of product is in FDA event 89054?
The event has 2 product records. The first record gives this quantity: 187 Medical Device Cards
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2021 have the same class?
89% of the 900 device events that started in 2021 are in Class II.