Device recall: VITROS Chemistry Products Cl- Slides, Product Code 844 5207 Product Usage: For in vitro diagnostic use only. VITROS Chem
FDA enforcement event 80697 is a device recall by Ortho-Clinical Diagnostics of Rochester, NY, initiated 11 Jul 2018 and classified Class III, with 2 product records.
FDA defines Class III as a situation in which use of or exposure to a violative product is not likely to cause adverse health consequences. 4% of the 1,074 device enforcement events initiated in 2018 carry the same class.
Status: Terminated, terminated 11 Aug 2020. Initiation: voluntary: firm initiated. Ortho-Clinical Diagnostics has 105 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Ortho-Clinical Diagnostics, Rochester, NY.
Reason for recall
"The slides do not meet the maximum allowable interference (MAI) claims for triglyceride interference when using serum samples. Triglyceride interference on the Cl- Slide was observed around a triglyceride concentration 600mg/dL, however the current triglyceride interference claim states interference should not occur below 800mg/dL of triglyceride."
Distribution
"Worldwide Distribution - US Nationwide and in the countries of: Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Columbia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway,…"
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-3043-2018 | VITROS Chemistry Products Cl- Slides, Product Code 844 5207 Product Usage: For in vitro diagnostic use only. VITROS Chemistry Products Cl- Slides quantitatively measure chloride (Cl-) concentration in serum and plasma using VITROS 250/350/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System. | 128196 | Class III | Terminated | 19 Sep 2018 |
| Z-3044-2018 | VITROS Chemistry Products Cl- Slides, Product Code 684 4471 Product Usage For in vitro diagnostic use only. VITROS Chemistry Products Cl- Slides quantitatively measure chloride (Cl-) concentration in serum, plasma and urine using VITROS 250/350/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.… | 420105 | Class III | Terminated | 19 Sep 2018 |
Context
Other enforcement events for Ortho-Clinical Diagnostics
All 105 events for Ortho-Clinical Diagnostics
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "The slides do not meet the maximum allowable interference (MAI) claims for triglyceride interference when using serum samples. Triglyceride interference on the Cl- Slide was observed around a triglyceride concentration 600mg/dL, however the current triglyceride interference claim states interference should not occur below 800mg/dL of triglyceride."
What does Class III mean?
FDA classifies recalls by relative degree of health hazard. Class III is a situation in which use of or exposure to a violative product is not likely to cause adverse health consequences.
Where was the product distributed?
FDA's distribution pattern field reads: "Worldwide Distribution - US Nationwide and in the countries of: Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Columbia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway,…"
Is the recall still ongoing?
FDA lists the status as Terminated; the termination date is 11 Aug 2020. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.