Recall Register

Food, drugs, devices › Ortho-Clinical Diagnostics

Device recall: VITROS Chemistry Products Calibrator Kit 20-in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 20- U

FDA enforcement event 92440 · Class II · initiated 8 May 2023 · status Ongoing

DeviceVoluntary: Firm initiated1 product recordRochester, NY

Ortho-Clinical Diagnostics (Rochester, NY) initiated this device recall on 8 May 2023; FDA lists 1 product record under event 92440, classified Class II.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 88% of the 917 device enforcement events initiated in 2023 carry the same class.

Status: Ongoing. Initiation: voluntary: firm initiated. Ortho-Clinical Diagnostics has 105 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Ortho-Clinical Diagnostics, Rochester, NY.

Reason for recall

"Potential to cause biased results in the upper end of the reportable range for the quantitative measurement of Transferrin, C3, C4, IgA, IgG and IgM. Within the reference interval, the highest bias observed in selected sample testing investigation was +35.3% for IgM; -4.3% for IgG; +12.3% for IgA; -6.9% for C3; -6.6% for C4; and -17.8% for transferrin. Falsely elevated or reduced results could…"

Distribution

"Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada L3R 4G5, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico,…"

Product records

Recall numberProductQuantityClassStatusReport date
Z-2129-2023VITROS Chemistry Products Calibrator Kit 20-in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 20- Used to calibrate the VITROS 5,1 FS/4600 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems for the quantitative measurement of transferrin, C3, C4, IgA, IgG and IgM. Product Code:…2400 unitsClass IIOngoing19 Jul 2023

Context

88%of 2023 device events are Class II917 events that year
1product record in this eventfirm median: 1
105events on this register for the firm

Other enforcement events for Ortho-Clinical Diagnostics

InitiatedProduct (first record)TypeClassRecords
27 Feb 2024VITROS Performance Verifier IDeviceClass II2
5 Jun 2023VITROS Immunodiagnostic Products Troponin I ES Reagent PackDeviceClass II2
20 Jul 2020VITROS XT3400 Chemistry System, Product code 6844458DeviceClass II2
17 Jul 2020VITROS Anti-SARS-CoV-2 Total Reagent, COV2TOT, VITROS CoV2T. Catalog number 6199922, UDI 10758750033386 - Product Usage:DeviceClass II2
13 Jul 2020VITROS Chemistry Products TBIL Slides- 5 PACK/300 SLDS In vitro Diagnostic quantitatively measure total bilirubin (TBIL)DeviceClass II3
20 May 2020VITROS 3600 Immunodiagnostic System Software Versions 3.3.3 and below Product Code: 6802783, Unique Identifier: 10758750DeviceClass II5
11 May 2020VITROS Chemistry Product AMYL Slides Reagent, AMYLASE/5 PACK/300 SLDS, Product Code 1202670, UDI # 10758750008513 - ProdDeviceClass II4
1 Apr 2020VITROS Chemistry Products CKMB Slides, Creatine Kinase MB, Product Code 805 8232-300 slides, UDI Number 10758750004294 -DeviceClass II2
10 Mar 2020VITROS Immunodiagnostic Products TT4 Reagent Pack, Product Code 8744468DeviceClass III1
10 Sep 2019VITROS chemistry Products Cl- Slides, Catalog No. 6844471 (which supports Urine CL-). MicroSlides which contain reagentsDeviceClass II3

All 105 events for Ortho-Clinical Diagnostics

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Potential to cause biased results in the upper end of the reportable range for the quantitative measurement of Transferrin, C3, C4, IgA, IgG and IgM. Within the reference interval, the highest bias observed in selected sample testing investigation was +35.3% for IgM; -4.3% for IgG; +12.3% for IgA; -6.9% for C3; -6.6% for C4; and -17.8% for transferrin. Falsely elevated or reduced results could…"

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada L3R 4G5, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico,…"

Is the recall still ongoing?

FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.