Food, drugs, devices › Ortho-Clinical Diagnostics
Device recall: VITROS Chemistry Products FS Calibrator 1, REF 680 1873 12 x 3 mL, Rx ONLY, IVD; used in conjunction with VITROS Chemist
Ortho-Clinical Diagnostics (Rochester, NY) initiated this device recall on 5 Jan 2017; FDA lists 1 product record under event 76200, classified Class III.
FDA defines Class III as a situation in which use of or exposure to a violative product is not likely to cause adverse health consequences. 4% of the 1,091 device enforcement events initiated in 2017 carry the same class.
Status: Terminated, terminated 22 Jun 2018. Initiation: voluntary: firm initiated. Ortho-Clinical Diagnostics has 105 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Ortho-Clinical Diagnostics, Rochester, NY.
Reason for recall
"The company received a customer complaint for the inability to calibrate VITROS Chemistry Products dLDL Reagent when using VITROS FS Calibrator 1, Lot 91461 and ADD (DRV 5908). Assay Data Disk (ADD) DRV versions 5904, 5905, 5906, 5907, 5908, 5909, and 5910 are all affected by this issue."
Distribution
"Worldwide Distribution - US (Nationwide and, Puerto Rico) and Foreign distribution to the following countries: Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India,…"
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1181-2017 | VITROS Chemistry Products FS Calibrator 1, REF 680 1873 12 x 3 mL, Rx ONLY, IVD; used in conjunction with VITROS Chemistry Products Calibrator Kits to calibrate VITROS 5,1 FS/4600 Chemistry and integrated Systems. --- Ortho-Clinical Diagnostics, Inc. Rochester, NY 14626 --- Ortho-Clinical Diagnostics Pencoed, CF35… | US: 160 kits, Foreign: 642 kits | Class III | Terminated | 15 Feb 2017 |
Context
Other enforcement events for Ortho-Clinical Diagnostics
All 105 events for Ortho-Clinical Diagnostics
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "The company received a customer complaint for the inability to calibrate VITROS Chemistry Products dLDL Reagent when using VITROS FS Calibrator 1, Lot 91461 and ADD (DRV 5908). Assay Data Disk (ADD) DRV versions 5904, 5905, 5906, 5907, 5908, 5909, and 5910 are all affected by this issue."
What does Class III mean?
FDA classifies recalls by relative degree of health hazard. Class III is a situation in which use of or exposure to a violative product is not likely to cause adverse health consequences.
Where was the product distributed?
FDA's distribution pattern field reads: "Worldwide Distribution - US (Nationwide and, Puerto Rico) and Foreign distribution to the following countries: Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India,…"
Is the recall still ongoing?
FDA lists the status as Terminated; the termination date is 22 Jun 2018. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.