Recall Register

Device recall: VITROS CRP Slides: 1) VITROS Chemistry Products CRP Slides (250 slides per sales unit), Catalog # 1926740 and 2)…

Class III FDA enforcement event 75692. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

CRP samples diluted using VITROS Chemistry Products Specialty Diluent, VITROS Chemistry Products FS Diluent Pack 3, or a sample with a low concentration of CRP had positively biased results if the endogenous level of CRP in the diluent is not subtracted from the final result. The IFUs do not provide specific instructions for how to account for endogenous CRP in the diluent when calculating the…

Class III, in the words of FDA

a situation in which use of or exposure to a violative product is not likely to cause adverse health consequences

Ortho-Clinical Diagnostics started this device recall on 26 Oct 2016. The firm address in the FDA record is in Rochester, NY. FDA put the recall in Class III. FDA event 75692 has 1 product record. The first FDA enforcement report date is 28 Dec 2016. The record gives the status Terminated, with the termination date 22 Jun 2018. 317 of the 1,232 device events that started in 2016 have more product records. The median device event of 2016 has 1 product record. The openFDA file has 1,232 device events that started in 2016 and 1,153 that started in 2015.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico, and to the countries of : Bermuda, Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Republic of…

Recalling firm

Ortho-Clinical Diagnostics, Rochester, NY.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 75692: product records
Recall numberProductQuantityClassStatusReport date
Z-0864-2017VITROS CRP Slides: 1) VITROS Chemistry Products CRP Slides (250 slides per sales unit), Catalog # 1926740 and 2), VITROS Chemistry Products CRP Slides (90 slides per sales unit), Catalog # 1926740124,063 unitsClass IIITerminated28 Dec 2016

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class III: The lowest of the three FDA recall classes.

The event in context

Device events that started in 2016
1,232
Share of these events in Class III
5%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
105

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 75692. https://recallregister.com/fda/iii/75692/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
10 Nov 2026
Newest FDA report date on the register
4 Nov 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico, and to the countries of : Bermuda, Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Republic of…
What quantity of product is in FDA event 75692?
The FDA record gives this quantity: 124,063 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2016 have the same class?
5% of the 1,232 device events that started in 2016 are in Class III.