Device recall: Liquichek" ToRCH Plus IgM Control - Positive Liquichek ToRCH Plus IgM Control is intended for use as an unassayed…
Class III FDA enforcement event 64959. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
The Liquichek" ToRCH Plus IgM Control - Positive, contains detectable levels of HCV RNA which originated from the non-specific human IgM component used in the mouse monoclonal antibody conjugation process for this product.
Class III, in the words of FDA
a situation in which use of or exposure to a violative product is not likely to cause adverse health consequences
Bio-rad Laboratories, Inc. started this device recall on 15 Apr 2013. The firm address in the FDA record is in Irvine, CA. FDA put the recall in Class III. FDA event 64959 has 1 product record. The first FDA enforcement report date is 22 May 2013. The record gives the status Terminated, with the termination date 1 Aug 2013.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide Distribution - Distributed in San Juan Puerto Rico and the states of NM, NY, MN, OR and WI., and the countries of Australia, Brazil, France, Italy, Korea, and New Zealand.
Recalling firm
Bio-rad Laboratories, Inc., Irvine, CA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1330-2013 | Liquichek" ToRCH Plus IgM Control - Positive Liquichek ToRCH Plus IgM Control is intended for use as an unassayed quality control serum to monitor precision of IgM laboratory testing procedures for the analytes listed in this package insert. This product is not intended for use in blood donor screening assays.… | 211 boxes and 5 MiniPaks | Class III | Terminated | 22 May 2013 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class III: The lowest of the three FDA recall classes.
The event in context
- Device events that started in 2013
- 1,236
- Share of these events in Class III
- 5%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 20
Other events of the firm
Official channels
- FDA enforcement report Z-1330-2013The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 64959. https://recallregister.com/fda/iii/64959/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide Distribution - Distributed in San Juan Puerto Rico and the states of NM, NY, MN, OR and WI., and the countries of Australia, Brazil, France, Italy, Korea, and New Zealand.
- What quantity of product is in FDA event 64959?
- The FDA record gives this quantity:
211 boxes and 5 MiniPaks
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2013 have the same class?
- 5% of the 1,236 device events that started in 2013 are in Class III.