Recall Register

Food, drugs, devices › Bio-Rad Laboratories, Inc.

Device recall: BioPlex 2200, Anti-CCP Reagent Pack, REF 665-3250, IVD

FDA enforcement event 91854 · Class II · initiated 21 Feb 2023 · status Ongoing

DeviceVoluntary: Firm initiated2 product recordsRedmond, WA

Bio-Rad Laboratories, Inc. (Redmond, WA) initiated this device recall on 21 Feb 2023; FDA lists 2 product records under event 91854, classified Class II.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 88% of the 917 device enforcement events initiated in 2023 carry the same class.

Status: Ongoing. Initiation: voluntary: firm initiated. Bio-Rad Laboratories, Inc. has 20 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Bio-Rad Laboratories, Inc., Redmond, WA.

Reason for recall

"Bio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8  96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 …"

Distribution

"Worldwide distribution - US Nationwide and the countries of Canada, Czech Republic, Germany, Spain, France, United Kingdom, Israel, Italy, Norway, Sweden."

Product records

Recall numberProductQuantityClassStatusReport date
Z-1377-2023BioPlex 2200, Anti-CCP Reagent Pack, REF 665-3250, IVD1544 unitsClass IIOngoing19 Apr 2023
Z-1378-2023BioPlex 2200, Anti-CCP Calibrator Set, REF 663-3200, IVD225 unitsClass IIOngoing19 Apr 2023

Context

88%of 2023 device events are Class II917 events that year
2product records in this eventfirm median: 1
20events on this register for the firm

Other enforcement events for Bio-Rad Laboratories, Inc.

InitiatedProduct (first record)TypeClassRecords
30 May 2025Platelia Toxo IgM Catalog Number 26211 UDI Code: 03610520005552 The Platelia Toxo IgM assay is a semi-quantitative immunDeviceClass II1
28 Feb 2023BioPlex 2200 REF 665-2050, APLS IgM Pack, APLS IgM Reagent PackDeviceClass II1
19 Oct 2022CMV IgM EIA, in vitro diagnostic.DeviceClass II1
14 Apr 2022BioPlex 2200 REF 12000670 ToRC IgM Reagent PackDeviceClass II1
7 Apr 2022Resin UPDATE CD-ROM, REF: 250-3020 inside VARIANTnbs Sickle Cell Program Reorder Pack, REF: 250-3000DeviceClass II1
8 Feb 2022BioPlex 2200 REF 12000650 SYPH T PACK Syphilis Total & RPR Reagent PackDeviceClass II2
17 Sep 2021BIO-RAD BioPlex 2200 REF 12000650 SYPH T PACK Syphilis Total & RPR Reagent Pack The BioPlex Syphilis Total & RPR kit is DeviceClass II2
21 Apr 2020BioPlex 2200 Syphilis Total & RPR Linked Immunoabsorption Assay, Treponema pallidum:UDI:03610520653012 - Product Usage: DeviceClass II1
1 Nov 2019BioPlex 2200 ANA Screen Control Set, REF: 663-1131 incompatible with ANA Screen Reagent Packs/ANA Kits, REF: 665-1150; aDeviceClass II1
10 Jan 2019EVOLIS Microplate System (Catalog no. 89700) with TimeLiner Data Management Software, (Catalog no. 90471A.DeviceClass II1

All 20 events for Bio-Rad Laboratories, Inc.

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Bio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8  96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 …"

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide distribution - US Nationwide and the countries of Canada, Czech Republic, Germany, Spain, France, United Kingdom, Israel, Italy, Norway, Sweden."

Is the recall still ongoing?

FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.