Device recall: Medline Renewal LigaSure Exact, Dissector without Nano-Coating 21cm FT10 SW Version 4.0.1.15 or lower; Electrosurgical…
Class II FDA enforcement event 97822. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
Medline Renewal has determined that certain areas on the device have the potential to trap residual material, which may remain within the device and, in rare cases, become dislodged.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Medline Industries, LP started this device recall on 1 Oct 2025. The firm address in the FDA record is in Northfield, IL. FDA put the recall in Class II. FDA event 97822 has 11 product records. The first FDA enforcement report date is 19 Nov 2025. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
US Nationwide distribution.
Recalling firm
Medline Industries, LP, Northfield, IL.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
11 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0467-2026 | Medline Renewal LigaSure Exact, Dissector without Nano-Coating 21cm FT10 SW Version 4.0.1.15 or lower; Electrosurgical Cutting And Coagulation Device | 4226 units | Class II | Ongoing | 19 Nov 2025 |
| Z-0468-2026 | Medline Renewal LigaSure Impact Curved Large Jaw Sealer/Divider, Compatible with FT10 Generator, FT10 SW Version 4.0.1.15 or lower; Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic, Reprocessed | 1 unit | Class II | Ongoing | 19 Nov 2025 |
| Z-0469-2026 | Medline Renewal Ligasure Impact Sealer/Divider NanoCoated, Compatible with FT10 Generator 18cm FT10 SW Version 4.0.1.15 or lower; Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic, Reprocessed | 58,499 units | Class II | Ongoing | 19 Nov 2025 |
| Z-0470-2026 | Medline Renewal LigaSure Small Jaw Instrument (Purple/White), Compatible with FT10 Generator 18.8 cm FT10 SW Version 4.0.1.15 or lower; Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic, Reprocessed | 18,117 units | Class II | Ongoing | 19 Nov 2025 |
| Z-0471-2026 | Medline Renewal LigaSure Blunt Tip Sealer/Divider, Compatible with FT10 Generator, 5 mm x 37 cm FT10 SW Version 4.0.1.15 or lower; Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic, Reprocessed | 2 units | Class II | Ongoing | 19 Nov 2025 |
| Z-0472-2026 | Medline RenewalLigasure Blunt Tip Sealer/Divider, NanoCoated Compatible with FT10 Generator 23 cm FT10 SW Version 4.0.1.15 or lower; Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic, Reprocessed | 65 units | Class II | Ongoing | 19 Nov 2025 |
| Z-0473-2026 | Medline Renewal Ligasure Blunt Tip Sealer/Divider NanoCoated, Compatible with FT10 Generator 37 cm FT10 SW Version 4.0.1.15 or lower; Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic, Reprocessed | 60,917 units | Class II | Ongoing | 19 Nov 2025 |
| Z-0474-2026 | Medline Renewal Ligasure Blunt Tip Sealer/Divider NanoCoated, Compatible with FT10 Generator 44cm FT10 SW Version 4.0.1.15 or lower; Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic, Reprocessed | 1,446 units | Class II | Ongoing | 19 Nov 2025 |
| Z-0475-2026 | Medline Renewal LigaSure Maryland Jaw Sealer/Divider Nano Coated, Compatible with FT10 Generator 23 cm FT10 SW Version 4.0.1.15 or lower; Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic, Reprocessed | 4,007 units | Class II | Ongoing | 19 Nov 2025 |
| Z-0476-2026 | Medline Renewal LigaSure Maryland Jaw Sealer/Divider NanoCoated, Compatible w/FT10 37 cm FT10 SW Version 4.0.1.15 or lower; Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic, Reprocessed | 57,244 units | Class II | Ongoing | 19 Nov 2025 |
| Z-0477-2026 | Medline Renewal Ligasure Maryland Jaw Sealer/Divider NanoCoated, Compatible with FT10 Generator 44 cm FT10 SW Version 4.0.1.15 or lower; Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic, Reprocessed | 9,547 units | Class II | Ongoing | 19 Nov 2025 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2025
- 945
- Share of these events in Class II
- 85%
- Product records in this event
- 11Median of the firm: 5
- Events of the firm on the register
- 66
Other events of the firm
Official channels
- FDA enforcement report Z-0467-2026The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 97822. https://recallregister.com/fda/ii/97822/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
US Nationwide distribution.
- What quantity of product is in FDA event 97822?
- The event has 11 product records. The first record gives this quantity:
4226 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2025 have the same class?
- 85% of the 945 device events that started in 2025 are in Class II.