Recall Register

Food, drugs, devices › Medline Industries, LP

Device recall: Medline medical convenience kits, labeled as: 1) SINUS PACK-LF, Medline Kit SKU DYNJ0966146F; 2) ENT PACK, Medline Kit S

FDA enforcement event 99489 · Class II · initiated 8 Jul 2026 · status Ongoing

DeviceVoluntary: Firm initiated4 product recordsNorthfield, IL

Medline Industries, LP (Northfield, IL) initiated this device recall on 8 Jul 2026; FDA lists 4 product records under event 99489, classified Class II.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 87% of the 572 device enforcement events initiated in 2026 carry the same class.

Status: Ongoing. Initiation: voluntary: firm initiated. Medline Industries, LP has 66 enforcement events on this register, with a median of 5 product records per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Medline Industries, LP, Northfield, IL.

Reason for recall

"Specific item(s) and lot(s) of Medline convenience kits containing Pilling Wecksorb Neurosponges and Neurosponge Strips are being recalled due to potential out-of-specification levels of endotoxin of the neurosponges."

Distribution

"US Nationwide distribution in the states of AZ, CA, IA, ID, IL, IN, MA, MD, MI, MN, MO, MT, NY, OH, TX, WI, WY."

Product records

Recall numberProductQuantityClassStatusReport date
Z-2994-2026Medline medical convenience kits, labeled as: 1) SINUS PACK-LF, Medline Kit SKU DYNJ0966146F; 2) ENT PACK, Medline Kit SKU DYNJ40154D; 3) ENT PACK, Medline Kit SKU DYNJ40154F; 4) ENT SCOPE PACK, Medline Kit SKU DYNJ44480D; 5) ENT NASAL ORAL PACK MOSES, Medline Kit SKU DYNJ58928A; 6) HEAD & NECK PACK, Medline Kit SKU…2262 kitsClass IIOngoing2 Sep 2026
Z-2995-2026Medline medical convenience kits, labeled as: 1) CRANIOTOMY PACK, Medline Kit SKU DYNJT6545; 2) ORTHO SPINE PACK, Medline Kit SKU DYNJ88056A; 3) ENT SINUS PACK, Medline Kit SKU DYNJ89001; 4) THYROID ENT, Medline Kit SKU DYNJ908641.246 kitsClass IIOngoing2 Sep 2026
Z-2996-2026Medline medical convenience kits, labeled as: 1) NEURO CDS, Medline Kit SKU CDS985246J; 2) NEURO PACK, Medline Kit SKU DYNJ32770V; 3) NEURO BUNDLE-LF, Medline Kit SKU DYNJ40095D; 4) NEURO PACK-LF, Medline Kit SKU DYNJ42844D; 5) NEURO PACK, Medline Kit SKU DYNJ45121I; 6) NEURO LAMINECTOMY PACK, Medline Kit SKU…3468 kitsClass IIOngoing2 Sep 2026
Z-2997-2026Medline medical convenience kits, labeled as: 1) LAMINECTOMY PACK, Medline Kit SKU DYNJT6548; 2) ORTHO SPINE PACK, Medline Kit SKU DYNJ44504G; 3) NEURO SPINE PACK, Medline Kit SKU DYNJ45419F; 4) CRANIOTOMY PACK, Medline Kit SKU DYNJ46024F; 5) SPINE PACK, Medline Kit SKU DYNJ54479C; 6) CRANIOTOMY PACK, Medline Kit SKU…974 kitsClass IIOngoing2 Sep 2026

Context

87%of 2026 device events are Class II572 events that year
4product records in this eventfirm median: 5
66events on this register for the firm

Other enforcement events for Medline Industries, LP

InitiatedProduct (first record)TypeClassRecords
11 Aug 2026Medline Convenience Kits: 1) KIT ANESTHESIA L&D A- LINE SET, Catalog Number: DYKMBNDL199ADeviceClass I7
31 Jul 2026Convenience kits ANESTHESIA PAIN PK LEAGUE DYNJ67925ADeviceClass II4
23 Jul 2026Medline Convenience Kit: OPEN HEART A PACK, Product Number DYNJ43008NDeviceClass II1
20 Jul 2026Medline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators PAIN PRE OP KIT DYKS1476ADeviceClass II13
13 Jul 2026Medline Hudson RCI, DBD-CIRCUIT,VENT,INSPIRATORY,HEATED,NEO, Medline Product Number/SKU HUD78019 (Teleflex Product NumbeDeviceClass II3
10 Jul 2026Medline medical convenience kit labeled as: TRACH TUBE KIT, Medline Kit SKU DYKM2330DeviceClass II1
10 Jul 2026Medline convenience kits labeled as: 1) FEMORAL PACK, Medline Kit SKU DYNJ909393A; 2) FEVAR, Medline Kit SKU DYNJ910295DDeviceClass II11
30 Jun 2026Medline Micro-Kill Bleach Germicidal Bleach Wipes,7" x 8", 150 count, 150 count, Product Number MSC351400ANDeviceClass II4
12 Jun 2026Medline Convenience Kits: 1) DRAWER 9 CRICOTHYROTOMY KIT, Model Number: ACC010741; 2) KIT ANESTHESIA L&D A- LINE SET, MoDeviceClass II16
11 Jun 2026Medline medical convenience kits labeled as: 1) KIT PEDIATRIC OPEN HEART MOSES, Medline Kit SKU DYNJ910103B; 2) BASIC HEDeviceClass II4

All 66 events for Medline Industries, LP

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Specific item(s) and lot(s) of Medline convenience kits containing Pilling Wecksorb Neurosponges and Neurosponge Strips are being recalled due to potential out-of-specification levels of endotoxin of the neurosponges."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "US Nationwide distribution in the states of AZ, CA, IA, ID, IL, IN, MA, MD, MI, MN, MO, MT, NY, OH, TX, WI, WY."

Is the recall still ongoing?

FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance