Recall Register

Device recall: Carescape B450, Model no. 5805686 - shipped with potentially affected batteries or affected batteries installed as a…

Class II FDA enforcement event 97786. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

GE Healthcare Finland Oy started this device recall on 26 Sep 2025. The firm address in the FDA record is in Helsinki. FDA put the recall in Class II. FDA event 97786 has 5 product records. The first FDA enforcement report date is 19 Nov 2025. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide - US Nationwide distribution.

Recalling firm

GE Healthcare Finland Oy, Helsinki, Finland.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

5 product records in the event.

FDA event 97786: product records
Recall numberProductQuantityClassStatusReport date
Z-0453-2026Carescape B450, Model no. 5805686 - shipped with potentially affected batteries or affected batteries installed as a spare part Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms Record updated 3/19/26 to include: CARESCAPE B450 MBA313 2095800-001 CARESCAPE MONITOR B450 (ATO MODEL) 2068491-002…3271Class IIOngoing19 Nov 2025
Z-0454-2026CARESCAPE ONE, Model no. 2087075-300 - shipped with potentially affected batteries or affected batteries installed as a spare part Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms Updated 3/19 to add CARESCAPE ONE MBZ101 2087075-0015935Class IIOngoing19 Nov 2025
Z-0455-2026B105P, B125P, B105M, B125M, B155M Patient Monitors; Model nos. 6160000-002 6160000-003 6160000-004 6160000-005 6160000-101 6160000-102 6160000-103 6160000-104 6160000-105 6160001-063 6160001-064 6160001-065 Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms 3/19/26 updated to add B105P Patient…7408Class IIOngoing19 Nov 2025
Z-0456-2026Portrait VSM Model nos. 6660000-666 Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms5990Class IIOngoing19 Nov 2025
Z-0457-2026BATTERY FLEX-3S2P 10.8V 18650 LI-ION SMBUS used with affected patient monitors Model No 2062895-001 Manufacturer Part Number U80296-1R01, revision S2857Class IIOngoing19 Nov 2025

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2025
945
Share of these events in Class II
85%
Product records in this event
5Median of the firm: 5
Events of the firm on the register
5

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 97786. https://recallregister.com/fda/ii/97786/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide - US Nationwide distribution.
What quantity of product is in FDA event 97786?
The event has 5 product records. The first record gives this quantity: 3271
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2025 have the same class?
85% of the 945 device events that started in 2025 are in Class II.