Device recall: CARESCAPE Canvas 1000, Model Numbers: a) 5865770; b) 5865770-01040169; c) 5865770-01054142; d) 5865770-01054696; e)…
Class II FDA enforcement event 94494. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
GE HealthCare has become aware of the potential for loss of monitoring on CARESCAPE Canvas 1000, CARESCAPE Canvas Smart Display, and certain versions of CARESCAPE B850 and CARESCAPE B650 monitors if, following the replacement of the CPU timekeeper battery, a mains power loss occurs during active monitoring. In the unlikely event that this occurs, it can result in the displays or monitors not…
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
GE Healthcare Finland Oy started this device recall on 5 Apr 2024. The firm address in the FDA record is in Helsinki. FDA put the recall in Class II. FDA event 94494 has 5 product records. The first FDA enforcement report date is 3 Jul 2024. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide distribution.
Recalling firm
GE Healthcare Finland Oy, Helsinki, Finland.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
5 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2168-2024 | CARESCAPE Canvas 1000, Model Numbers: a) 5865770; b) 5865770-01040169; c) 5865770-01054142; d) 5865770-01054696; e) 5865770-01054698; f) 5865770-01054721; g) 5865770-01054740; h) 5865770-01055202; i) 5865770-01055204; j) 5865770-01055270; k) 5865770-01055317; l) 5865770-01055796; m) 5865770-01058501; n)… | 9267 units | Class II | Ongoing | 3 Jul 2024 |
| Z-2169-2024 | CARESCAPE Canvas Smart Display, Model Numbers: a) 5700023, b) 5700023-02, c) 5939590-102; monitor, physiological, patient | 4649 units | Class II | Ongoing | 3 Jul 2024 |
| Z-2170-2024 | CARESCAPE Canvas Smart Display, Model Numbers: a) 5514418, b) 5514418 / MAEP2, c) 5514418-01059642, d) 5514418-01080402, e) 5514418-01081317, f) 5514418-01086007, g) 5514418-01086838, h) 5514418-01088182, i) 5514418-01108442, j) 5514418-01118673, k) 5514418-01120764, l) 5514418-01166521, m) 5514418-01180241, n)… | 547 units | Class II | Ongoing | 3 Jul 2024 |
| Z-2171-2024 | CARESCAPE B650: Model Numbers: a) MBB313, Model Numbers: 1) 2095801-001, 2) 2095801-001-01018416, 3) 2095801-001-01023150, 4) 2095801-001-01023151, 5) 2095801-001-01023152, 6) 2095801-001-01032962, 7) 2095801-001-01033915, 8) 2095801-001-01034302, 9) 2095801-001-01034399, 10) 2095801-001-01034528, 11)… | 54800 units | Class II | Ongoing | 3 Jul 2024 |
| Z-2172-2024 | CARESCAPE B850 Model Numbers: a) MBC303, Model Numbers: 1) 2078633-001, 2) 2078633-010, 3) 2078633-012, 4) 2078633-017, 5) 2078633-028, 6) 2095802-001-54069330, 7) 2095802-001-554922, 8) 2095802-001-554923, 9) 2095802-001-554924, 10) 2095802-001-554925, 11) 2095802-001-554926, 12) 2095802-001-554927, 13)… | 25268 units | Class II | Ongoing | 3 Jul 2024 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2024
- 970
- Share of these events in Class II
- 87%
- Product records in this event
- 5Median of the firm: 5
- Events of the firm on the register
- 5
Other events of the firm
| Initiated | Product (first record) | Type | Class | Records |
|---|---|---|---|---|
| 26 Sep 2025 | Carescape B450, Model no. 5805686 - shipped with potentially affected batteries or affected batteries installed as a spa | Device | Class II | 5 |
| 19 May 2023 | TruSignal Adult/Pediatric Sensor; Oximeter, Ear: a) REF TS-AP-10, 10 pcs; b) REF TS-AP-25, 25 pcs | Device | Class I | 9 |
| 4 Sep 2018 | GE Healthcare CARESCAPE Monitor B450 Product Usage: Intended uses of CARESCAPE B450 in 510(k) countries [UR_CRSCP-885]: | Device | Class II | 2 |
| 25 May 2018 | GE Healthcare CARESCAPE Monitor B650 | Device | Class II | 1 |
Official channels
- FDA enforcement report Z-2168-2024The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 94494. https://recallregister.com/fda/ii/94494/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide distribution.
- What quantity of product is in FDA event 94494?
- The event has 5 product records. The first record gives this quantity:
9267 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2024 have the same class?
- 87% of the 970 device events that started in 2024 are in Class II.