Device recall: Philips Allura Xper Interventional Fluoroscopic X-ray Systems, specifically the Disk Bay of the PC component of…
Class II FDA enforcement event 93781. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
Disk Bay may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. started this device recall on 2 Jan 2024. The firm address in the FDA record is in Best. FDA put the recall in Class II. FDA event 93781 has 2 product records. The first FDA enforcement report date is 28 Feb 2024. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide Distribution: US nationwide including states of: AL, AK, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NV, NY, OH,…
Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., Best, N/A, Netherlands.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
2 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1151-2024 | Philips Allura Xper Interventional Fluoroscopic X-ray Systems, specifically the Disk Bay of the PC component of affected systems. Affected PCs: Imaging Processing PC, Host PC, FlexVision PC Allura Xper Model Name - Model Number: Allura Xper FD10 722003 722010 722026; Allura Xper FD10 OR Table 722022; Allura Xper… | 13, 133 (3,408 US; 9,725 ROW) | Class II | Ongoing | 28 Feb 2024 |
| Z-1152-2024 | Philips Azurion Interventional Fluoroscopic X-ray Systems, specifically the Disk Bay of the PC component of the affected system. Affected PCs: XRay PC, Suite PC, and FlexViewing PC Azurion Model Name - Model Number: Azurion 3 M12 722063 722221; Azurion 3 M15 722064 722222; Azurion 5 M20 722228; Azurion 7 B12 722067… | 13, 133 (3,408 US; 9,725 ROW) | Class II | Ongoing | 28 Feb 2024 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2024
- 970
- Share of these events in Class II
- 87%
- Product records in this event
- 2Median of the firm: 1
- Events of the firm on the register
- 59
Other events of the firm
Official channels
- FDA enforcement report Z-1151-2024The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 93781. https://recallregister.com/fda/ii/93781/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide Distribution: US nationwide including states of: AL, AK, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NV, NY, OH,…
- What quantity of product is in FDA event 93781?
- The event has 2 product records. The first record gives this quantity:
13, 133 (3,408 US; 9,725 ROW)
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2024 have the same class?
- 87% of the 970 device events that started in 2024 are in Class II.