Device recall: Trilogy Evo and Trilogy Evo O2 Material Numbers BL2110X15B BR2110X18B CA2110X12B DE2100X13B DE2110X13B DS2100X11B…
Class II FDA enforcement event 93609. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
The following was missing from the Contraindications Statement: The AVAPS-AE therapy mode is contraindicated for invasive use and patients less than 10 kg. The AVAPS feature is contraindicated for patients less than 10 kg. Potential harms if devices are used on contraindicated patients include barotrauma, hypoventilation/hypercapnia, and rebreathing of excessive CO2.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Philips Respironics, Inc. started this device recall on 27 Nov 2023. The firm address in the FDA record is in Murrysville, PA. FDA put the recall in Class II. FDA event 93609 has 1 product record. The first FDA enforcement report date is 10 Jan 2024. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide distribution: US (nationwide)-AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC,…
Recalling firm
Philips Respironics, Inc., Murrysville, PA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0638-2024 | Trilogy Evo and Trilogy Evo O2 Material Numbers BL2110X15B BR2110X18B CA2110X12B DE2100X13B DE2110X13B DS2100X11B DS2110X11B EE2100X15B EE2110X15B ES2100X15B ES2110X15B EU2100X15B EU2100X19 EU2110X15B FR2100X14B FR2110X14B FX2100X15B GB2110X15B IA2100X15B IA2110X15B IN2100X15B IN2100X19 IN2110X15B IT2100X21B… | 100,126 (54,691 US; 45,435 OUS) in total | Class II | Ongoing | 10 Jan 2024 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2023
- 917
- Share of these events in Class II
- 88%
- Product records in this event
- 1Median of the firm: 3
- Events of the firm on the register
- 26
Other events of the firm
Official channels
- FDA enforcement report Z-0638-2024The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 93609. https://recallregister.com/fda/ii/93609/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide distribution: US (nationwide)-AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC,…
- What quantity of product is in FDA event 93609?
- The FDA record gives this quantity:
100,126 (54,691 US; 45,435 OUS) in total
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2023 have the same class?
- 88% of the 917 device events that started in 2023 are in Class II.