Recall Register

Food, drugs, devices › Philips Respironics, Inc.

Device recall: Trilogy Evo O2. Model Numbers: FP2100X10, DS2100X11B, IN2100X15B, DE2100X13B, IA2100X15B, EU2100X15B, LA2100X15B, EE2100

FDA enforcement event 95486 · Class I · initiated 7 Oct 2024 · status Ongoing

DeviceVoluntary: Firm initiated7 product recordsMurrysville, PA

FDA enforcement event 95486 is a device recall by Philips Respironics, Inc. of Murrysville, PA, initiated 7 Oct 2024 and classified Class I, with 7 product records.

FDA defines Class I as a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death. 11% of the 970 device enforcement events initiated in 2024 carry the same class.

Status: Ongoing. Initiation: voluntary: firm initiated. Philips Respironics, Inc. has 26 enforcement events on this register, with a median of 3 product records per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Philips Respironics, Inc., Murrysville, PA.

Reason for recall

"Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements."

Distribution

"Worldwide distribution - US Nationwide and the countries of AE, AF, AR, AT, AU, BD, BE, BG, BH, BO, BR, BS, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GE, GF, GI, GR, GT, HK,…"

Product records

Recall numberProductQuantityClassStatusReport date
Z-0290-2025Trilogy Evo O2. Model Numbers: FP2100X10, DS2100X11B, IN2100X15B, DE2100X13B, IA2100X15B, EU2100X15B, LA2100X15B, EE2100X15B, FX2100X15B, ES2100X15B, ND2100X15B, FR2100X14B, IT2100X21B, JP2100X16B, IN2100X19, EU2100X19. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at…24,249 unitsClass IOngoing20 Nov 2024
Z-0291-2025Trilogy Evo. Model Numbers: DS2110X11B, UDS2110X11B, RDS2110X11B, CA2110X12B, JP2110X16B, IN2110X15B, DE2110X13B, AU2110X15B, BL2110X15B, EU2110X15B, BR2110X18B, LA2110X15B, EE2110X15B, FR2110X14B, GB2110X15B, ND2110X15B, ES2110X15B, IA2110X15B, IT2110X21B, KR2110X15B, TR2110X15B, EU2110X19, FP2110X10, PP2110X10,…82,800 unitsClass IOngoing20 Nov 2024
Z-0292-2025Trilogy Evo Universal Ventilator. Model Number DS2000X11B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings.9,999 unitsClass IOngoing20 Nov 2024
Z-0293-2025Trilogy EV300. Model Numbers: DS2200X11B, IN2200X15B, BL2200X15B, BR2200X18B, CA2200X12B, DE2200X13B, GB2200X15B, ND2200X15B, ES2200X15B, FR2200X14B, FX2200X15B, EU2200X15B, IA2200X15B, IT2200X21B, KR2200X15B, TR2200X15B, EE2200X15B. Intended for invasive and non-invasive ventilation of pediatric through adult…34,146 unitsClass IOngoing20 Nov 2024
Z-0294-2025Aeris EVO. Model Number: VT2110X24B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings.804 unitsClass IOngoing20 Nov 2024
Z-0295-2025Garbin Evo. Model Number: LD2110X23B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings.1,322 unitsClass IOngoing20 Nov 2024
Z-0296-2025LifeVent EVO2. Model Number: SP2100X26B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings.1,244 unitsClass IOngoing20 Nov 2024

Context

11%of 2024 device events are Class I970 events that year
7product records in this eventfirm median: 3
26events on this register for the firm

Other enforcement events for Philips Respironics, Inc.

InitiatedProduct (first record)TypeClassRecords
2 Mar 2026Philips Respironics Trilogy Evo, Software Version 1.05.15.00. Continuous home-use ventilator device.DeviceClass I4
2 Mar 2026Philips Respironics Trilogy Evo, Software Version 1.05.15.00. Continuous home-use ventilator device.DeviceClass I4
7 Oct 2025SimplyGo Mini, Standard Battery Kit; Product code: 1116816; The affected battery kits are sold as accessories for the SiDeviceClass II1
30 Jun 2025DreamStation Auto CPAP. Non-Continuous Ventilator.DeviceClass I3
9 May 2025BiPAP V30 Auto Ventilator. Intended to provide non-invasive ventilatory support for pediatric and adult patients weighinDeviceClass I3
16 Jul 2024Trilogy Evo EV300 -- Model Number/ Model Description: BL2200X15B Trilogy Evo, O2, Benelux EV300; BR2200X18B Trilogy Evo,DeviceClass I3
1 Apr 2024OmniLab Advanced + Ventilator Product Numbers 1111122 1111123 1111124 1111125 1111126 1111127 1111138 1111141 1111142 KRDeviceClass I1
26 Mar 2024BiPAP V30 Auto Ventilator, Part Number 1111178DeviceClass I3
19 Mar 2024Aeris EVO, all software versions except 1.05.06.00. Ventilator for pediatric and adult patients. (sold outside US)DeviceClass I7
27 Nov 2023Trilogy Evo and Trilogy Evo O2 Material Numbers BL2110X15B BR2110X18B CA2110X12B DE2100X13B DE2110X13B DS2100X11B DS2110DeviceClass II1

All 26 events for Philips Respironics, Inc.

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements."

What does Class I mean?

FDA classifies recalls by relative degree of health hazard. Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide distribution - US Nationwide and the countries of AE, AF, AR, AT, AU, BD, BE, BG, BH, BO, BR, BS, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GE, GF, GI, GR, GT, HK,…"

Is the recall still ongoing?

FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.