Food, drugs, devices › Philips Respironics, Inc.
Device recall: Trilogy Evo O2. Model Numbers: FP2100X10, DS2100X11B, IN2100X15B, DE2100X13B, IA2100X15B, EU2100X15B, LA2100X15B, EE2100
FDA enforcement event 95486 is a device recall by Philips Respironics, Inc. of Murrysville, PA, initiated 7 Oct 2024 and classified Class I, with 7 product records.
FDA defines Class I as a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death. 11% of the 970 device enforcement events initiated in 2024 carry the same class.
Status: Ongoing. Initiation: voluntary: firm initiated. Philips Respironics, Inc. has 26 enforcement events on this register, with a median of 3 product records per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Philips Respironics, Inc., Murrysville, PA.
Reason for recall
"Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements."
Distribution
"Worldwide distribution - US Nationwide and the countries of AE, AF, AR, AT, AU, BD, BE, BG, BH, BO, BR, BS, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GE, GF, GI, GR, GT, HK,…"
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0290-2025 | Trilogy Evo O2. Model Numbers: FP2100X10, DS2100X11B, IN2100X15B, DE2100X13B, IA2100X15B, EU2100X15B, LA2100X15B, EE2100X15B, FX2100X15B, ES2100X15B, ND2100X15B, FR2100X14B, IT2100X21B, JP2100X16B, IN2100X19, EU2100X19. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at… | 24,249 units | Class I | Ongoing | 20 Nov 2024 |
| Z-0291-2025 | Trilogy Evo. Model Numbers: DS2110X11B, UDS2110X11B, RDS2110X11B, CA2110X12B, JP2110X16B, IN2110X15B, DE2110X13B, AU2110X15B, BL2110X15B, EU2110X15B, BR2110X18B, LA2110X15B, EE2110X15B, FR2110X14B, GB2110X15B, ND2110X15B, ES2110X15B, IA2110X15B, IT2110X21B, KR2110X15B, TR2110X15B, EU2110X19, FP2110X10, PP2110X10,… | 82,800 units | Class I | Ongoing | 20 Nov 2024 |
| Z-0292-2025 | Trilogy Evo Universal Ventilator. Model Number DS2000X11B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings. | 9,999 units | Class I | Ongoing | 20 Nov 2024 |
| Z-0293-2025 | Trilogy EV300. Model Numbers: DS2200X11B, IN2200X15B, BL2200X15B, BR2200X18B, CA2200X12B, DE2200X13B, GB2200X15B, ND2200X15B, ES2200X15B, FR2200X14B, FX2200X15B, EU2200X15B, IA2200X15B, IT2200X21B, KR2200X15B, TR2200X15B, EE2200X15B. Intended for invasive and non-invasive ventilation of pediatric through adult… | 34,146 units | Class I | Ongoing | 20 Nov 2024 |
| Z-0294-2025 | Aeris EVO. Model Number: VT2110X24B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings. | 804 units | Class I | Ongoing | 20 Nov 2024 |
| Z-0295-2025 | Garbin Evo. Model Number: LD2110X23B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings. | 1,322 units | Class I | Ongoing | 20 Nov 2024 |
| Z-0296-2025 | LifeVent EVO2. Model Number: SP2100X26B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings. | 1,244 units | Class I | Ongoing | 20 Nov 2024 |
Context
Other enforcement events for Philips Respironics, Inc.
All 26 events for Philips Respironics, Inc.
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements."
What does Class I mean?
FDA classifies recalls by relative degree of health hazard. Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.
Where was the product distributed?
FDA's distribution pattern field reads: "Worldwide distribution - US Nationwide and the countries of AE, AF, AR, AT, AU, BD, BE, BG, BH, BO, BR, BS, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GE, GF, GI, GR, GT, HK,…"
Is the recall still ongoing?
FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.
Where does this data come from?
openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.
Provenance
Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.