Device recall: Patient Information Center iX with Software Version Number 4.x in the following: (1) Patient Information Center iX…
Class II FDA enforcement event 92992. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
Patient Information Center (PIC iX) Release 4.x Surveillance Crash Caused by Intel Graphics Driver Error-with a blank screen and subsequently require a manual reboot of the hardware to restart and continue central patient monitoring. Potential for a delay in the detection of a change or deterioration in the condition of one or more patients
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Philips North America Llc started this device recall on 17 Aug 2023. The firm address in the FDA record is in Cambridge, MA. FDA put the recall in Class II. FDA event 92992 has 1 product record. The first FDA enforcement report date is 27 Sep 2023. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Nationwide Foreign: Austria Bahrain Belgium Canada Czech Republic Denmark Finland France Germany Greece India Ireland Italy Japan Jordan Kenya Kuwait Lebanon Lithuania Netherlands Norway Poland…
Recalling firm
Philips North America Llc, Cambridge, MA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2608-2023 | Patient Information Center iX with Software Version Number 4.x in the following: (1) Patient Information Center iX Product Number: 866389; (2) Patient Information Center iX Expand Product Number: 866390; (3) PIC iX Essentials Product Number: 867093; (4) PIC iX Essentials Expand Product Number: 867154 | 960 units | Class II | Ongoing | 27 Sep 2023 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2023
- 917
- Share of these events in Class II
- 88%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 129
Other events of the firm
Official channels
- FDA enforcement report Z-2608-2023The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 92992. https://recallregister.com/fda/ii/92992/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Nationwide Foreign: Austria Bahrain Belgium Canada Czech Republic Denmark Finland France Germany Greece India Ireland Italy Japan Jordan Kenya Kuwait Lebanon Lithuania Netherlands Norway Poland…
- What quantity of product is in FDA event 92992?
- The FDA record gives this quantity:
960 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2023 have the same class?
- 88% of the 917 device events that started in 2023 are in Class II.