Recall Register

Food, drugs, devices › Covidien, LP

Device recall: Step Bladeless Trocars, 10 mm cannula and dilator with radially expandable sleeve and 5 - 10 mm reducer, Product Number

FDA enforcement event 91873 · Class II · initiated 1 Mar 2023 · status Ongoing

DeviceVoluntary: Firm initiated4 product recordsNorth Haven, CT

FDA enforcement event 91873 is a device recall by Covidien, LP of North Haven, CT, initiated 1 Mar 2023 and classified Class II, with 4 product records.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 88% of the 917 device enforcement events initiated in 2023 carry the same class.

Status: Ongoing. Initiation: voluntary: firm initiated. Covidien, LP has 31 enforcement events on this register, with a median of 2 product records per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Covidien, LP, North Haven, CT.

Reason for recall

"Medtronic was made aware of a supplier calibration issue during the contract sterilization process that may lead to unsterile product. Use of affected product may lead to an unspecified infection or, in the case of sutures used in the eye, pose a loss of vision risk."

Distribution

"Worldwide Distribution: US (nationwide) to States of: AR, CA, FL, GA, LA, MA, NC, NY, OH, WA; and countries of: Canada, Colombia, El Salvador, Indonesia, Nepal, Puerto Rico, Taiwan, and Thailand."

Product records

Recall numberProductQuantityClassStatusReport date
Z-1387-2023Step Bladeless Trocars, 10 mm cannula and dilator with radially expandable sleeve and 5 - 10 mm reducer, Product Number S101010770 unitsClass IIOngoing19 Apr 2023
Z-1388-2023Chromic Gut Absorbable Sutures, SURGIGUT* 0 CHR GS-25 90CM X36 (CG-904), SURGIGUT* 1 CHR GS-24 90CM X36 (CG-915), CHROMIC GUT 4-0 75CM V20 (GG-121), CHROMIC GUT 3-0 75CM V20 X36 (GG-122), CHROMIC GUT 2-0 75CM V20 X36 (GG-123), CHROMIC GUT 2-0 75CM V26 (GG-127), CHROMIC GUT 4-0 45CM P12 X36 (SG-5637), CHROMIC GUT 5-0…36204 unitsClass IIOngoing19 Apr 2023
Z-1389-2023Mild Chromic Gut Absorbable Sutures, CH GUT 5-0 12 UD SS-14 SS-14 (G-1766-K), CHROMIC GUT 6-0 45CM HE-1 X12 (G-1780K)960 unitsClass IIOngoing19 Apr 2023
Z-1390-2023Plain Gut Absorbable Sutures, PLAIN GUT 5/0 18 P-13 X12 (SG-5686), PLAIN GUT 4-0 18 UD SC-11 DA (SG828-2)3228 unitsClass IIOngoing19 Apr 2023

Context

88%of 2023 device events are Class II917 events that year
4product records in this eventfirm median: 2
31events on this register for the firm

Other enforcement events for Covidien, LP

InitiatedProduct (first record)TypeClassRecords
24 Jun 2026V-Loc 90 Absorbable Wound Closure Device VLOC 90 0 VL 22 GS-21 VLOCM0336 VLOC 90 0 VL 22 GS-25 VLOCM0436 VLOCM0436 VLOCMDeviceClass II5
15 Apr 2026Covidien EndoStitch" Polysorb Single Use Loading Unit; Product numbers: 170053 and 170071;DeviceClass II1
25 Sep 2025Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); CFN: SIGSDL45CTVT;DeviceClass II2
19 Sep 2024Chameleon PTA Balloon Catheter with Injection Port 5mm x 40mm x 75cm (CFN CH05-40-75US), 6mm x 40mm x 75cm (CFN CH06-40-DeviceClass II1
15 Apr 2024Tri-Staple 2.0 Black Reinforced Intelligent Reload- Applications in abdominal, gynecologic, pediatric and thoracic surgeDeviceClass II1
15 Mar 2024Palindrome Precision HSI Chronic Catheter Sport Pack 14.5 Fr/Ch (4.8mm) x 19 cm (Symmetrical Tip, Heparin Coating, SilveDeviceClass II2
28 Feb 2024Covidien Auto Suture" Structural Balloon Trocar, Product Number OMS-T10SBDeviceClass II7
14 Feb 2024Signia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVTDeviceClass II2
25 Jan 2024Sofsilk Braided Silk sutures, Product Number S-2782K. Nonabsorbable, sterile, non-mutagenic surgical suture composed of DeviceClass II1
21 Dec 2023Surgilon Braided Nylon sutures: 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT 88861919-41 SURGILON* 3-0 BLK 7X75CMPCT 88861919DeviceClass II2

All 31 events for Covidien, LP

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Medtronic was made aware of a supplier calibration issue during the contract sterilization process that may lead to unsterile product. Use of affected product may lead to an unspecified infection or, in the case of sutures used in the eye, pose a loss of vision risk."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide Distribution: US (nationwide) to States of: AR, CA, FL, GA, LA, MA, NC, NY, OH, WA; and countries of: Canada, Colombia, El Salvador, Indonesia, Nepal, Puerto Rico, Taiwan, and Thailand."

Is the recall still ongoing?

FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.