Recall Register

Device recall: Surgilon Braided Nylon sutures: 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT 88861919-41 SURGILON* 3-0 BLK 7X75CMPCT…

Class II FDA enforcement event 93768. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Specific lots of sutures were sterilized with gamma doses that exceeded the range approved. The extra gamma levels may decrease the strength of the sutures over time, which could result in harms such as wound dehiscence and hemorrhage/blood loss/ bleeding at a critical level.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Covidien LP started this device recall on 21 Dec 2023. The firm address in the FDA record is in Mansfield, MA. FDA put the recall in Class II. FDA event 93768 has 2 product records. The first FDA enforcement report date is 7 Feb 2024. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide US Nationwide distribution including in the states of AL, AZ, CA, FL, GA, HI, IA, IL, IN, MD, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, OK, OR, PA, PR, TN, TX, VA, WA, WV and the countries…

Recalling firm

Covidien LP, Mansfield, MA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

2 product records in the event.

FDA event 93768: product records
Recall numberProductQuantityClassStatusReport date
Z-0844-2024Surgilon Braided Nylon sutures: 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT 88861919-41 SURGILON* 3-0 BLK 7X75CMPCT 88861919-51 SURGILON* 2-0 BLK 7X75CMPCT 88861919-71 SURGILON* 1 BLK 7X75CM PCT46,812 US; 68,976 OUSClass IIOngoing7 Feb 2024
Z-0845-2024Sofsilk Braided Silk sutures: CS-211 SOFSILK* 4-0 BLK 45CM CVF21 X36 GS-831 SOFSILK* 4-0 BLK 75CM V20 X36 S-176 SOFSILK* 0 BLK 12X45CM PCT X24 S-182 SOFSILK* 5-0 BLK 12X45CM PCT X36 S-184 SOFSILK* 3-0 BLK 12X45CM PCT X36 S-185 SOFSILK* 2-0 BLK 12X45CM PCT X36 S-245 SOFSILK* 2-0 BLK 12X60CM PCT X36 S-246 SOFSILK* 0…46,812 US; 68,976 OUSClass IIOngoing7 Feb 2024

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2023
917
Share of these events in Class II
88%
Product records in this event
2Median of the firm: 2
Events of the firm on the register
31

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 93768. https://recallregister.com/fda/ii/93768/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide US Nationwide distribution including in the states of AL, AZ, CA, FL, GA, HI, IA, IL, IN, MD, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, OK, OR, PA, PR, TN, TX, VA, WA, WV and the countries…
What quantity of product is in FDA event 93768?
The event has 2 product records. The first record gives this quantity: 46,812 US; 68,976 OUS
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2023 have the same class?
88% of the 917 device events that started in 2023 are in Class II.