Device recall: Short tibial bearing MK4 - Std, STM, Catalog No. SMMLTB02. Knee prosthesis component.
Class II FDA enforcement event 90836. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
A MK4 Tibial Bearing was supplied with the intention to mate with the Tibial Metal Casing in-situ. However, the device in-situ was a MK3 Tibial Metal Casing. The supplied MK4 tibial bearing was not exchangeable for the MK3 tibial bearing. The surgeon needed to complete the surgery by re-assembling and re-inserting the extracted MK3 tibial bearing.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Howmedica Osteonics Corp. started this device recall on 24 Aug 2022. The firm address in the FDA record is in Mahwah, NJ. FDA put the recall in Class II. FDA event 90836 has 1 product record. The first FDA enforcement report date is 5 Oct 2022. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Distributed to one customer located in New York.
Recalling firm
Howmedica Osteonics Corp., Mahwah, NJ.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1778-2022 | Short tibial bearing MK4 - Std, STM, Catalog No. SMMLTB02. Knee prosthesis component. | 1 unit | Class II | Ongoing | 5 Oct 2022 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2022
- 839
- Share of these events in Class II
- 88%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 48
Other events of the firm
Official channels
- FDA enforcement report Z-1778-2022The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 90836. https://recallregister.com/fda/ii/90836/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Distributed to one customer located in New York.
- What quantity of product is in FDA event 90836?
- The FDA record gives this quantity:
1 unit
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2022 have the same class?
- 88% of the 839 device events that started in 2022 are in Class II.