Device recall: TS Universal Baseplate Finishing Tray, Triathlon. Catalog Number: 6543-6-506.

Class II FDA enforcement event 99883. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Howmedica Osteonics Corp. started this device recall on 16 Sep 2026. The firm address in the FDA record is in Mahwah, NJ. FDA put the recall in Class II. FDA event 99883 has 4 product records. The first FDA enforcement report date is 21 Oct 2026. The record gives the status Ongoing. 94 of the 677 device events that started in 2026 have more product records. The median device event of 2026 has 1 product record.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.

Recalling firm

Howmedica Osteonics Corp., Mahwah, NJ.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

4 product records in the event.

FDA event 99883: product records
Recall numberProductQuantityClassStatusReport date
Z-0164-2027TS Universal Baseplate Finishing Tray, Triathlon. Catalog Number: 6543-6-506.605 traysClass IIOngoing21 Oct 2026
Z-0165-2027TS Femoral Prep Tray, Triathlon. Catalog Number: 6543-6-507.582 traysClass IIOngoing21 Oct 2026
Z-0166-2027Triathlon Hinge Femoral Prep Tray. Catalog Number: 6543-6-514.359 traysClass IIOngoing21 Oct 2026
Z-0167-2027TS Misc. Instruments Tray, Triathlon. Catalog Number: 6543-6-519.577 traysClass IIOngoing21 Oct 2026

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2026
677
Share of these events in Class II
87%
Product records in this event
4Median of the firm: 1
Events of the firm on the register
49

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 99883. https://recallregister.com/fda/ii/99883/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
27 Oct 2026
Newest FDA report date on the register
21 Oct 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.
What quantity of product is in FDA event 99883?
The event has 4 product records. The first record gives this quantity: 605 trays
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2026 have the same class?
87% of the 677 device events that started in 2026 are in Class II.