Recall Register

Device recall: VITROS Immunodiagnostic Products FSH Reagent Pack- IVD 1931922

Class II FDA enforcement event 90047. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Falsely high or delayed patient and QC results may occur due to low end imprecision. VITROS Immunodiagnostic Products FSH, LH and Prol Reagents Low End Imprecision. Low-level quality control (QC) and patient samples at the low end of the Measuring (Reportable) Range when using VITROS FSH, LH and Prol Reagent. Calibration failures, driven by imprecision observed with Calibrator Level 1.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Ortho-Clinical Diagnostics, Inc started this device recall on 25 Mar 2022. The firm address in the FDA record is in Raritan, NJ. FDA put the recall in Class II. FDA event 90047 has 3 product records. The first FDA enforcement report date is 25 May 2022. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution: US (Nationwide) OUS (Foreign): Australia Bermuda Brazil Canada L3R 4G5 Chile China Colombia Denmark France Germany India Italy Japan Mexico Netherlands Norway Portugal Russia…

Recalling firm

Ortho-Clinical Diagnostics, Inc, Raritan, NJ.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

3 product records in the event.

FDA event 90047: product records
Recall numberProductQuantityClassStatusReport date
Z-1112-2022VITROS Immunodiagnostic Products FSH Reagent Pack- IVD 19319222502 total: 2154 US; 348 OUSClass IIOngoing25 May 2022
Z-1113-2022VITROS Immunodiagnostic Products LH Reagent Pack Product Code: 13501981575 units total: 1183 US; 392 OUSClass IIOngoing25 May 2022
Z-1114-2022VITROS Immunodiagnostic Products Prolactin Reagent Pack Product Code: 18497931767 units total: 1521 US; 246 OUSClass IIOngoing25 May 2022

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2022
839
Share of these events in Class II
88%
Product records in this event
3Median of the firm: 1
Events of the firm on the register
54

Other events of the firm

Other FDA enforcement events of Ortho-Clinical Diagnostics, Inc
InitiatedProduct (first record)TypeClassRecords
29 Jul 2025Chemistry Products LAC Slides. Catalog Number: 8433880. 1 Sales Unit = contains 5 cartridges of 60 slides, for a total oDeviceClass II2
9 Jul 2025VITROS Chemistry Products Ca Slides; Catalog Number: 145 0261;DeviceClass II1
12 Jun 2025VITROS Chemistry Products OP Reagent Gen 15. Model Number: 680 1997. The VITROS OP Reagent is a dual-chambered package cDeviceClass II1
10 Jun 2025Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products Performance Verifier I Model/Catalog NumbeDeviceClass II4
9 May 2025Brand Name: VITROS Product Name: VITROS 5600 Integrated System Model/Catalog Number: (1) 6802413, (2) 6802915DeviceClass II3
14 Apr 2025VITROS Chemistry Products CRP Slides. One sales unit contains 5 cartridges, of 18 slides per cartridge, for a total of 9DeviceClass II1
24 Mar 2025Brand Name: Solana Bordetella Complete Assay Product Name: Solana Bordetella kit Model/Catalog Number: M308 Software VerDeviceClass II1
19 Dec 2024Brand Name: VITROS Product Name: VITROS 5600 Integrated System Model/Catalog Number: 6802413 (New), 6802415 (Certified/RDeviceClass II3
11 Oct 2024VITROS Chemistry Products CRP Slides- In vitro diagnostic quantitatively measures C-reactive protein (CRP) concentrationDeviceClass II1
10 Jul 2024The VITROS XT Chemistry Products ALB-TP Slides is a single device that contains both an albumin test and a total proteinDeviceClass II1

Ortho-Clinical Diagnostics, Inc: all 54 events

Official channels

Cite this page

Recall Register. FDA enforcement event 90047. https://recallregister.com/fda/ii/90047/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution: US (Nationwide) OUS (Foreign): Australia Bermuda Brazil Canada L3R 4G5 Chile China Colombia Denmark France Germany India Italy Japan Mexico Netherlands Norway Portugal Russia…
What quantity of product is in FDA event 90047?
The event has 3 product records. The first record gives this quantity: 2502 total: 2154 US; 348 OUS
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2022 have the same class?
88% of the 839 device events that started in 2022 are in Class II.