Recall Register

Food, drugs, devices › Ortho-Clinical Diagnostics, Inc.

Device recall: VITROS Chemistry Products OP Reagent Gen 15. Model Number: 680 1997. The VITROS OP Reagent is a dual-chambered package c

FDA enforcement event 97034 · Class II · initiated 12 Jun 2025 · status Ongoing

DeviceVoluntary: Firm initiated1 product recordRochester, NY

Ortho-Clinical Diagnostics, Inc. (Rochester, NY) initiated this device recall on 12 Jun 2025; FDA lists 1 product record under event 97034, classified Class II.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 85% of the 945 device enforcement events initiated in 2025 carry the same class.

Status: Ongoing. Initiation: voluntary: firm initiated. Ortho-Clinical Diagnostics, Inc. has 54 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Ortho-Clinical Diagnostics, Inc., Rochester, NY.

Reason for recall

"Lots from Generation (GEN) 15 of VITROS Chemistry Products OP Reagent may generate lower than expected quality control (QC) results. If the customer is unable to get passing quality control results, then the customer would be unable to run the OP-LO protocol leading to a potential delay in patient results."

Distribution

"Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, France, Germany, India, Italy, Japan, Mexico, Norway,…"

Product records

Recall numberProductQuantityClassStatusReport date
Z-2073-2025VITROS Chemistry Products OP Reagent Gen 15. Model Number: 680 1997. The VITROS OP Reagent is a dual-chambered package containing ready-to-use liquid reagents that are used to detect opiates in urine. One sales unit contains 6 Microtip packs of reagent. Each pack has 50 OP tests.2,017 unitsClass IIOngoing9 Jul 2025

Context

85%of 2025 device events are Class II945 events that year
1product record in this eventfirm median: 1
54events on this register for the firm

Other enforcement events for Ortho-Clinical Diagnostics, Inc.

InitiatedProduct (first record)TypeClassRecords
29 Jul 2025Chemistry Products LAC Slides. Catalog Number: 8433880. 1 Sales Unit = contains 5 cartridges of 60 slides, for a total oDeviceClass II2
9 Jul 2025VITROS Chemistry Products Ca Slides; Catalog Number: 145 0261;DeviceClass II1
10 Jun 2025Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products Performance Verifier I Model/Catalog NumbeDeviceClass II4
9 May 2025Brand Name: VITROS Product Name: VITROS 5600 Integrated System Model/Catalog Number: (1) 6802413, (2) 6802915DeviceClass II3
14 Apr 2025VITROS Chemistry Products CRP Slides. One sales unit contains 5 cartridges, of 18 slides per cartridge, for a total of 9DeviceClass II1
24 Mar 2025Brand Name: Solana¿ Bordetella Complete Assay Product Name: Solana Bordetella kit Model/Catalog Number: M308 Software VeDeviceClass II1
19 Dec 2024Brand Name: VITROS Product Name: VITROS 5600 Integrated System Model/Catalog Number: 6802413 (New), 6802415 (Certified/RDeviceClass II3
11 Oct 2024VITROS Chemistry Products CRP Slides- In vitro diagnostic quantitatively measures C-reactive protein (CRP) concentrationDeviceClass II1
10 Jul 2024The VITROS XT Chemistry Products ALB-TP Slides is a single device that contains both an albumin test and a total proteinDeviceClass II1
3 Jul 2024VITROS Immunodiagnostic VITROS Folate Reagent Pack 1/2DeviceClass II1

All 54 events for Ortho-Clinical Diagnostics, Inc.

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Lots from Generation (GEN) 15 of VITROS Chemistry Products OP Reagent may generate lower than expected quality control (QC) results. If the customer is unable to get passing quality control results, then the customer would be unable to run the OP-LO protocol leading to a potential delay in patient results."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, France, Germany, India, Italy, Japan, Mexico, Norway,…"

Is the recall still ongoing?

FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.