Device recall: NeuMoDx Cartridge-For in vitro diagnostic use with NeuMoOx Molecular Systems with either of the following assays…
Class II FDA enforcement event 89846. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
Potential for false positive results for SARS-CoV-2 in identified lots of cartridges, when used in conjunction with assays containing SARS-CoV-2 targets (NeuMoDx" SARS-CoV-2 Test Strip Cat #300800 and NeuMoDx" Flu A-B/RSV/SARS-CoV-2 Vantage Assay Cat #300900) result incorrect management of patients
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
NeuMoDx Molecular Inc started this device recall on 18 Feb 2022. The firm address in the FDA record is in Ann Arbor, MI. FDA put the recall in Class II. FDA event 89846 has 1 product record. The first FDA enforcement report date is 4 May 2022. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide distribution - US Nationwide distribution in the states of OH, MO, VA, MI, MI, FL, MD, IA, NY, MO and the countries of Italy, Germany, Switzerland, Sweden, Netherlands, Finland, Spain,…
Recalling firm
NeuMoDx Molecular Inc, Ann Arbor, MI.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1000-2022 | NeuMoDx Cartridge-For in vitro diagnostic use with NeuMoOx Molecular Systems with either of the following assays: NeuMoDx SARS CoV-2 Test Strip (REF 300800) NeuMoDx Flu A-B/RSV/SARS-CoV-2 Vantage Test Strip (REF 300900) NeuMoDx Laboratory Developed Tests (LDTs) for SARS-CoV-2 RNA detection Note: Results for targets… | 773 (72 US and 701 ROW) cases (48 cartridges/case) | Class II | Ongoing | 4 May 2022 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2022
- 839
- Share of these events in Class II
- 88%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 6
Other events of the firm
| Initiated | Product (first record) | Type | Class | Records |
|---|---|---|---|---|
| 24 May 2023 | NeuMoDx Cartridge-used for extraction, purification, amplification and detection of nucleic acids on the NeuMoDx 288 and | Device | Class II | 1 |
| 15 May 2023 | NeuMoDx SARS-CoV-2 Assay -IVD intended for the qualitative detection of SARS-CoV-2 RNA from nasal, nasopharyngeal and or | Device | Class III | 1 |
| 13 May 2022 | NeuMoDx Flu A-B/RSV/SARS-CoV-2 Vantage Test Strip, Model No. 300900 | Device | Class II | 1 |
| 21 Jan 2022 | NeuMoDx Flu A-B/RSV/SARS-CoV-2 Vantage Test Strip, Ref # 300900 (marketed under EUA202947). Distributed in boxes of 6 as | Device | Class II | 1 |
| 4 May 2021 | NeuMoDx Cartridge, For In Vitro Diagnostic Use on the NeuMoDx 288 and NeuMoDx 96 Molecular Systems, Model Number 100100 | Device | Class II | 1 |
Official channels
- FDA enforcement report Z-1000-2022The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 89846. https://recallregister.com/fda/ii/89846/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide distribution - US Nationwide distribution in the states of OH, MO, VA, MI, MI, FL, MD, IA, NY, MO and the countries of Italy, Germany, Switzerland, Sweden, Netherlands, Finland, Spain,…
- What quantity of product is in FDA event 89846?
- The FDA record gives this quantity:
773 (72 US and 701 ROW) cases (48 cartridges/case)
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2022 have the same class?