Food, drugs, devices › NeuMoDx Molecular Inc
Device recall: NeuMoDx Flu A-B/RSV/SARS-CoV-2 Vantage Test Strip, Model No. 300900
NeuMoDx Molecular Inc (Ann Arbor, MI) initiated this device recall on 13 May 2022; FDA lists 1 product record under event 90275, classified Class II.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 88% of the 839 device enforcement events initiated in 2022 carry the same class.
Status: Ongoing. Initiation: voluntary: firm initiated. NeuMoDx Molecular Inc has 6 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
NeuMoDx Molecular Inc, Ann Arbor, MI.
Reason for recall
"There is a potential for reduced analytical sensitivity for samples in the 1x - 3x LOD range for Flu A, RSV, and SARS-CoV-2 targets. This may lead to false negative results, allowing ongoing transmission to others and delaying appropriate treatment."
Distribution
"Domestic distribution to FL, NY, MD, PA, VA, MI, TN, NM. Foreign distribution to AT, AU, BE, CH, DE, FI, FR, GB, GR, HK, ID, IT, LT, NL, SA, SE, and SI."
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1251-2022 | NeuMoDx Flu A-B/RSV/SARS-CoV-2 Vantage Test Strip, Model No. 300900 | 459 kits US; 1626 kits OUS | Class II | Ongoing | 15 Jun 2022 |
Context
Other enforcement events for NeuMoDx Molecular Inc
| Initiated | Product (first record) | Type | Class | Records |
|---|---|---|---|---|
| 24 May 2023 | NeuMoDx Cartridge-used for extraction, purification, amplification and detection of nucleic acids on the NeuMoDx 288 and | Device | Class II | 1 |
| 15 May 2023 | NeuMoDx SARS-CoV-2 Assay -IVD intended for the qualitative detection of SARS-CoV-2 RNA from nasal, nasopharyngeal and or | Device | Class III | 1 |
| 18 Feb 2022 | NeuMoDx Cartridge-For in vitro diagnostic use with NeuMoOx Molecular Systems with either of the following assays: NeuMoD | Device | Class II | 1 |
| 21 Jan 2022 | NeuMoDx Flu A-B/RSV/SARS-CoV-2 Vantage Test Strip, Ref # 300900 (marketed under EUA202947). Distributed in boxes of 6 as | Device | Class II | 1 |
| 4 May 2021 | NeuMoDx Cartridge, For In Vitro Diagnostic Use on the NeuMoDx 288 and NeuMoDx 96 Molecular Systems, Model Number 100100 | Device | Class II | 1 |
All 6 events for NeuMoDx Molecular Inc
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "There is a potential for reduced analytical sensitivity for samples in the 1x - 3x LOD range for Flu A, RSV, and SARS-CoV-2 targets. This may lead to false negative results, allowing ongoing transmission to others and delaying appropriate treatment."
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "Domestic distribution to FL, NY, MD, PA, VA, MI, TN, NM. Foreign distribution to AT, AU, BE, CH, DE, FI, FR, GB, GR, HK, ID, IT, LT, NL, SA, SE, and SI."
Is the recall still ongoing?
FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.
Where does this data come from?
openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.
Provenance
Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.