Reason for recall, in the words of FDA

Due to a software defect, the Philips StentBoost Live R2.0 might not process X-ray images of the current run and instead show an image processed in a previous run. The image shown could be from the same patient or from a different patient.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Philips North America Llc started this device recall on 2 Feb 2022. The firm address in the FDA record is in Cambridge, MA. FDA put the recall in Class II. FDA event 89787 has 1 product record. The first FDA enforcement report date is 13 Apr 2022. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Global Distribution. US Nationwide.

Recalling firm

Philips North America Llc, Cambridge, MA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 89787: product records
Recall numberProductQuantityClassStatusReport date
Z-0874-2022Philips Allura (Model Numbers 722010, 722012, 722026, 722027, 722028, 722029, 722035) and Philips Azurion (Model Numbers 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222, 722223, 722224, 722225, 722226, 722227, 722228) running StentBoost Live (Model Number 459801781661) Software Version Number: R2.0338Class IIOngoing13 Apr 2022

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2022
839
Share of these events in Class II
88%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
129

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 89787. https://recallregister.com/fda/ii/89787/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Global Distribution. US Nationwide.
What quantity of product is in FDA event 89787?
The FDA record gives this quantity: 338
How did the recall start?
The FDA record says: Voluntary: Firm initiated