Device recall: Zenition 50, Model #718096

Class II FDA enforcement event 89437. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Firm discovered that the wireless foot switch can suddenly stop responding due to a firmware issue. If the wireless foot switch fails, there could be a delay or interruption of procedure if the hand switch is not used or if the wired foot switch is not connected to the system.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. started this device recall on 2 Dec 2021. The firm address in the FDA record is in Best. FDA put the recall in Class II. FDA event 89437 has 3 product records. The first FDA enforcement report date is 23 Feb 2022. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Domestic distribution to AZ CA FL GA HI IL KS KY MA MD ME MI MN MS NC NE NJ NV NY OH PA TN TX VA WA WV. Worldwide foreign distribution.

Recalling firm

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., Best, N/A, Netherlands.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

3 product records in the event.

FDA event 89437: product records
Recall numberProductQuantityClassStatusReport date
Z-0647-2022Zenition 50, Model #71809616 US; 76 ROWClass IIOngoing23 Feb 2022
Z-0648-2022Zenition 70, Model #71813395 US; 284 ROWClass IIOngoing23 Feb 2022
Z-0649-2022Veradius Unity, Model #71813212 US; 44 ROWClass IIOngoing23 Feb 2022

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2021
900
Share of these events in Class II
89%
Product records in this event
3Median of the firm: 1
Events of the firm on the register
59

Other events of the firm

Other FDA enforcement events of PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
InitiatedProduct (first record)TypeClassRecords
16 Jul 2026Philips Azurion Rystem, Model Numbers: 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222,722223, 722224, 72DeviceClass II1
16 Jul 2026Philips Azurion, Model Numbers: 722229, 722230, 722231, 722232, 722233, 722234, 722235, and 722236, with Software VersioDeviceClass II1
29 Jun 2026Allura Xper FD. Model Numbers: 722026, 722027, 722028, 722029, 722035, 722038, 722039, 722058. Interventional fluoroscopDeviceClass II1
17 Jun 2026Philips Allura under the following System Descriptions and corresponding model numbers: 1. Allura Xper FD10; Model NumbeDeviceClass II3
18 May 2026Philips MultiDiagnost Eleva with the following Model Numbers: Model # 708032 for MultiDiagnost Eleva; Model # 708034 forDeviceClass II2
30 Apr 2026Azurion system; System Code Description (Model Numbers): Azurion 3 M12 (722063, 722221), Azurion 3 M15 (722064, 722222),DeviceClass II2
28 Apr 2026Philips Azurion systems not configured with an optional auxiliary pan handle. Includes the below product descriptions anDeviceClass II1
28 Apr 2026Azurion system; System Code Description (Model Numbers): Azurion 3 M12 (722063), Azurion 3 M15 (722064), Azurion 7 B12 (DeviceClass II3
31 Mar 2026Philips Azurion. System Code Description (Model Number): Azurion 3 M12 (722221, 722229), Azurion 3 M15 (722222, 722230),DeviceClass II1
19 Mar 2026Philips AneurysmFlow. Model Number: 001015. AneurysmFlow is a software medical device (Interventional Tool) intended to DeviceClass II1

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: all 59 events

Official channels

Cite this page

Recall Register. FDA enforcement event 89437. https://recallregister.com/fda/ii/89437/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Domestic distribution to AZ CA FL GA HI IL KS KY MA MD ME MI MN MS NC NE NJ NV NY OH PA TN TX VA WA WV. Worldwide foreign distribution.
What quantity of product is in FDA event 89437?
The event has 3 product records. The first record gives this quantity: 16 US; 76 ROW
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2021 have the same class?
89% of the 900 device events that started in 2021 are in Class II.