Device recall: Trilogy Evo, Product numbers BL2110X15B, CA2110X12B, DE2110X13B, DS2110X11B, EE2110X15B, ES2110X15B, EU2110X15B…
Class II FDA enforcement event 88333. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
Two software issues have been identified related to pressure increase in the device: The first issue is described as "Infant/Pediatric EFS Calibration Pressure Increase" and the second issue is described as "Pressure Drift (Continuous Usage).
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Philips Respironics, Inc. started this device recall on 26 Jul 2021. The firm address in the FDA record is in Murrysville, PA. FDA put the recall in Class II. FDA event 88333 has 4 product records. The first FDA enforcement report date is 25 Aug 2021. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide distribution - US nationwide distribution.
Recalling firm
Philips Respironics, Inc., Murrysville, PA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
4 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2287-2021 | Trilogy Evo, Product numbers BL2110X15B, CA2110X12B, DE2110X13B, DS2110X11B, EE2110X15B, ES2110X15B, EU2110X15B, EU2110X19, FR2110X14B, GB2110X15B, IA2110X15B, IN2110X15B, IN2110X19, IT2110X21B, KR2110X15B, LA2110X15B, LD2110X23B, ND2110X15B, RDE2110X13B, UDS2110X11B | 18588 (US); 5662 (OUS) | Class II | Ongoing | 25 Aug 2021 |
| Z-2288-2021 | Trilogy Evo O2, Product numbers DE2100X13B, DS2100X11B, EE2100X15B, ES2100X15B, EU2100X15B, FR2100X14B, FX2100X15B, IA2100X15B, IN2100X15B, IN2100X19, IT2100X21B, JP2100X16B, LA2100X15B, ND2100X15B, RDE2100X13B, SP2100X26B | 101 (FEMA Stockpile Only); 21077 (OUS) | Class II | Ongoing | 25 Aug 2021 |
| Z-2289-2021 | Trilogy EV300, Product numbers CA2200X12B, DS2200X11B, FX2200X15B, IN2200X15B | 18936 (US); 4659 (OUS) | Class II | Ongoing | 25 Aug 2021 |
| Z-2290-2021 | Trilogy Evo Universal, Product number DS2000X11B | 10,000 (FEMA Stockpile Only) | Class II | Ongoing | 25 Aug 2021 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2021
- 900
- Share of these events in Class II
- 89%
- Product records in this event
- 4Median of the firm: 3
- Events of the firm on the register
- 26
Other events of the firm
Official channels
- FDA enforcement report Z-2287-2021The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 88333. https://recallregister.com/fda/ii/88333/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide distribution - US nationwide distribution.
- What quantity of product is in FDA event 88333?
- The event has 4 product records. The first record gives this quantity:
18588 (US); 5662 (OUS)
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2021 have the same class?
- 89% of the 900 device events that started in 2021 are in Class II.