Recall Register

Device recall: Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor Bracket Assembly for use with CT and PET/CT…

Class II FDA enforcement event 87974. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Philips North America Llc started this device recall on 1 May 2021. The firm address in the FDA record is in Cambridge, MA. FDA put the recall in Class II. FDA event 87974 has 12 product records. The first FDA enforcement report date is 7 Jul 2021. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution - US Nationwide and the countries of Canada, Mexico, AE, AR, AT, AU, BD, BE, BR, CH, CL, CN, CO, CU, CY, CZ, DE, DK, DZ, EE, EG, ES, FI, FR, GB, GE, GI, GL, GR, HK, HU, ID,…

Recalling firm

Philips North America Llc, Cambridge, MA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

12 product records in the event.

FDA event 87974: product records
Recall numberProductQuantityClassStatusReport date
Z-1979-2021Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor Bracket Assembly for use with CT and PET/CT systems: Model - MX8000 Models: 728120/728125/7282011,301 Monitor Bracket Assemblies totalClass IIOngoing7 Jul 2021
Z-1980-2021Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor Bracket Assembly for use with CT and MX16 Models: 7281311,301 Monitor Bracket Assemblies totalClass IIOngoing7 Jul 2021
Z-1981-2021Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor Bracket Assembly for use with CT and PET/CT systems. Brilliance 16 Power, Brilliance 16/10/6 Water/Air 728240/728245/728246 - Brilliance 16 728255/728256/728266 - Brilliance 6 728250/728251 - Brilliance 10 728260 - Extended Brilliance Workspace1,301 Monitor Bracket Assemblies totalClass IIOngoing7 Jul 2021
Z-1982-2021Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor Bracket Assembly for use with CT and PET/CT systems. Brilliance 64, Brilliance 40, Ingenuity Flex 728235 - Brilliance 40 728231/728232 - Brilliance 64 728317/728318 - Ingenuity Flex1,301 Monitor Bracket Assemblies totalClass IIOngoing7 Jul 2021
Z-1983-2021Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor Bracket Assembly for use with CT and PET/CT systems. BB (Big Bore) Oncology, BB (Big Bore) Radiology: 728242/728243/728244/728272 - Brilliance Big Bore1,301 Monitor Bracket Assemblies totalClass IIOngoing7 Jul 2021
Z-1984-2021Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor Ingenuity Core, Ingenuity Core 128, Ingenuity CT 728321 - Ingenuity Core 728323 - Ingenuity Core 128 728326/728327 - Ingenuity CT1,301 Monitor Bracket Assemblies totalClass IIOngoing7 Jul 2021
Z-1985-2021Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 728332 - IQon Spectral CT : IQon Spectral CT1,301 Monitor Bracket Assemblies totalClass IIOngoing7 Jul 2021
Z-1986-2021Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 728333 - Spectral CT:1,301 Monitor Bracket Assemblies totalClass IIOngoing7 Jul 2021
Z-1987-2021Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 882442/882456 - Ingenuity TF PET/CTN/AClass IIOngoing7 Jul 2021
Z-1988-2021Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 882446 - Vereos PET/CT1,301 Monitor Bracket Assemblies totalClass IIOngoing7 Jul 2021
Z-1989-2021Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 882476 - Gemini TF Big Bore PET/CT1,301 Monitor Bracket Assemblies totalClass IIOngoing7 Jul 2021
Z-1990-2021Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 728306 - Brilliance iCT 728311 - Brilliance iCT SP1,301 Monitor Bracket Assemblies totalClass IIOngoing7 Jul 2021

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2021
900
Share of these events in Class II
89%
Product records in this event
12Median of the firm: 1
Events of the firm on the register
129

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 87974. https://recallregister.com/fda/ii/87974/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution - US Nationwide and the countries of Canada, Mexico, AE, AR, AT, AU, BD, BE, BR, CH, CL, CN, CO, CU, CY, CZ, DE, DK, DZ, EE, EG, ES, FI, FR, GB, GE, GI, GL, GR, HK, HU, ID,…
What quantity of product is in FDA event 87974?
The event has 12 product records. The first record gives this quantity: 1,301 Monitor Bracket Assemblies total
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2021 have the same class?
89% of the 900 device events that started in 2021 are in Class II.